Evaluation of the i‐gel® Plus supraglottic airway device in elective surgery: a prospective international multicentre study*
Anaesthesia · 2024-12 · 79 (12) · 1284-1291
DOI: 10.1111/anae.16401 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Anaesthesia, online erschienen am 07.08.2024. Verfasst von Jakub Werner und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesia and Intensive Medicine, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic. Hauptthemen laut MEDLINE: Elective Surgical Procedures, Laryngeal Masks, Intubation, Intratracheal. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Ministerstvo Zdravotnictví Ceské Republiky.
Abstract (Originalsprache)
Background
The i‐gel® Plus is a modified version of the i‐gel® supraglottic airway device. It contains a wider drainage port; a longer tip; ramps inside the breathing channel; and an additional port for oxygen delivery. There has been no prospective evaluation of this device in clinical practice.
Methods
This international, multicentre, prospective cohort study aimed to evaluate the performance of the i‐gel Plus in adult patients undergoing elective procedures under general anaesthesia. The primary outcome was overall insertion success rate, defined as the ability to provide effective airway management through the device from insertion until the end of the surgical procedure. Secondary outcomes included device performance and incidence of postoperative adverse events. Data from the first 1000 patients are reported.
Results
In total, 1012 patients were enrolled; 12 forms were excluded from the final analysis due to incomplete data leaving 1000 included patients (545 female). Overall insertion success rate was 98.6%, with a first‐attempt success rate of insertion of 88.2%. A significant difference between females and males was seen for the overall success rate (97.4% vs. 99.6% respectively) but not for first‐attempt successful insertion. Mean (SD) oropharyngeal seal pressure was 32 (7) cmH 2 O. The only independent factor that increased the risk of first‐attempt failure was low operator experience. Complications included desaturation < 85% in 0.6%; traces of blood on the device in 7.4%; laryngospasm in 0.5%; and gastric contents inside the bowl in 0.2% of patients.
Conclusions
The i‐gel Plus appears to be an effective supraglottic airway device that is associated with a high insertion success rate and a reasonably low incidence of complications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anaesthesia and Intensive Medicine, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic.
- Masaryk University, Brno, Czech Republic.
- Department of Anaesthesiology and Intensive Care Medicine, University Hospital Olomouc, Olomouc, Czech Republic.
- Department of Anaesthesiology and Intensive Medicine, General University Hospital and First Medical Faculty of the Charles University, Prague, Czech Republic.
- Department of Anaesthesiology and Intensive Therapy, Medical University of Lodz, Lodz, Poland.
- Department of Anaesthesia and Intensive Medicine, Military University Hospital, Prague, Czech Republic.
ORCID-Profile: Pavel Michalek
Förderung
- Ministerstvo Zdravotnictví Ceské Republiky
Zitieren
APA
Werner, J., Klementova, O., Bruthans, J., Macoun, J., Gaszynski, T., Henlin, T., Donaldson, W., Lichnovsky, E., Arava, S., Lopez, A. M., Berge, R., & Michalek, P. (2024). Evaluation of the i‐gel® Plus supraglottic airway device in elective surgery: a prospective international multicentre study*. Anaesthesia, 79(12), 1284-1291. https://doi.org/10.1111/anae.16401
Vancouver
Werner J, Klementova O, Bruthans J, Macoun J, Gaszynski T, Henlin T, et al. Evaluation of the i‐gel® Plus supraglottic airway device in elective surgery: a prospective international multicentre study*. Anaesthesia. 2024;79(12):1284-1291. doi:10.1111/anae.16401
BibTeX
@article{Werner2024,
author = {Werner, Jakub and Klementova, Olga and Bruthans, Jan and Macoun, Jaromir and Gaszynski, Tomasz and Henlin, Tomas and Donaldson, Will and Lichnovsky, Erik and Arava, Shiva and Lopez, Ana M and Berge, Raquel and Michalek, Pavel},
title = {Evaluation of the i‐gel® Plus supraglottic airway device in elective surgery: a prospective international multicentre study*},
journal = {Anaesthesia},
year = {2024},
volume = {79},
number = {12},
pages = {1284--1291},
doi = {10.1111/anae.16401}
}