C‐Arm Cone Beam CT ‐Guided Preoperative Microcoil Pulmonary Ground Glass Nodule Localization: Diagnostic and Surgical Advantage
Thoracic Cancer · 2025
DOI: 10.1111/1759-7714.70152 ↗Kurz gefasst
Fachartikel in der Zeitschrift Thoracic Cancer, 2025. Verfasst von Carlo Altomare und 11 weiteren Autor:innen. Beteiligte Einrichtung: Unit of Diagnostic Imaging and Interventional Radiology, Fondazione Policlinico Universitario Campus Bio-Medico, Rome, Italy. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Thoracic Surgery, Video-Assisted, Lung Neoplasms, Solitary Pulmonary Nodule. Die Arbeit nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
This study evaluates the effectiveness and safety of C‐arm cone beam CT (CBCT)‐guided microcoil localization combined with uniportal video‐assisted thoracoscopic surgery (VATS) for the management of small, difficult‐to‐localize ground‐glass opacities (GGOs) and sub‐solid nodules in the lungs.
Methods
We retrospectively analyzed data from 13 patients with single, small, peripheral, non‐subpleural GGOs or SSN. All patients underwent successful microcoil localization using CB‐CT guidance followed by uniportal VATS resection. A microcoil was positioned partly in the lung parenchyma and partly in the extra‐pleural space to assist in intraoperative localization. We evaluated the rate of correct microcoil placement and the technical success of the resection.
Results
Microcoil placement was successfully performed in all patients, with an average procedure time of 28.8 ± 10.8 min. The mean nodule size was 9.9 ± 5.4 mm, and 76.9% of the nodules were classified as ground‐glass opacities. No intraparenchymal bleeding was observed, and four patients (30.8%) experienced pneumothorax, all of which were self‐limited and required no intervention or coil repositioning. The uniVATS resection success rate was 100%.
Conclusion
CBCT‐guided microcoil localization, with partial placement of the coil in the extra‐pleural space, proved to be a highly effective technique for the localization and resection of small pulmonary nodules. The procedure demonstrated high accuracy, minimal complications, reduction of procedural time, and short hospital stays. Intraoperative fluoroscopy was never necessary, with a high reduction in radiation exposure for the patient and the operator. Further studies with larger populations and longer follow‐ups are needed to validate these findings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Unit of Diagnostic Imaging and Interventional Radiology, Fondazione Policlinico Universitario Campus Bio-Medico, Rome, Italy.
- Unit of Diagnostic Imaging and Interventional Radiology, Department of Medicine and Surgery, Università Campus Bio-Medico di Roma, Rome, Italy.
- Department of Occupational and Environmental Medicine, Epidemiology and Hygiene, National Institute for Insurance Against Accidents at Work, Rome, Italy.
- Unit of Gastroenterology, Università Campus Bio-Medico di Roma, Rome, Italy.
- Unit of Thoracic Surgery, Fondazione Policlinico Universitario Campus Bio-Medico, Rome, Italy.
- Unit of Thoracic Surgery, Department of Medicine and Surgery, Università Campus Bio-Medico di Roma, Rome, Italy.
ORCID-Profile: Carlo Altomare
Zitieren
APA
Altomare, C., Casati, R., Pacella, G., Olivieri, L., Tirabasso, A., Altomare, A., Frasca, L., Longo, F., Crucitti, P., Faiella, E., Beomonte Zobel, B., & Grasso, R. F. (2025). C‐Arm Cone Beam CT ‐Guided Preoperative Microcoil Pulmonary Ground Glass Nodule Localization: Diagnostic and Surgical Advantage. Thoracic Cancer. https://doi.org/10.1111/1759-7714.70152
Vancouver
Altomare C, Casati R, Pacella G, Olivieri L, Tirabasso A, Altomare A, et al. C‐Arm Cone Beam CT ‐Guided Preoperative Microcoil Pulmonary Ground Glass Nodule Localization: Diagnostic and Surgical Advantage. Thoracic Cancer. 2025. doi:10.1111/1759-7714.70152
BibTeX
@article{Altomare2025,
author = {Altomare, Carlo and Casati, Rebecca and Pacella, Giuseppina and Olivieri, Laura and Tirabasso, Angelo and Altomare, Annamaria and Frasca, Luca and Longo, Filippo and Crucitti, Pierfilippo and Faiella, Eliodoro and Beomonte Zobel, Bruno and Grasso, Rosario Francesco},
title = {C‐Arm Cone Beam CT ‐Guided Preoperative Microcoil Pulmonary Ground Glass Nodule Localization: Diagnostic and Surgical Advantage},
journal = {Thoracic Cancer},
year = {2025},
doi = {10.1111/1759-7714.70152}
}
EUVIMED