Smart, Individualized Green Hemodiafiltration ‐ Effects of Individualized Dialysate Flow Through AutoFlow on Dialysate Consumption and Parameters of Dialysis Adequacy in the Setting of Online Hemodiafiltration

Vanessa David, Bastian Au, Torsten M Meyer, Kijanosh Lehmann iD, Jan T Kielstein iD

Therapeutic Apheresis and Dialysis · 2025

DOI: 10.1111/1744-9987.70072 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Therapeutic Apheresis and Dialysis, 2025. Verfasst von Vanessa David und 4 weiteren Autor:innen. Beteiligte Einrichtung: Medical Clinic V, Nephrology, Rheumatology, and Blood Purification, Academic Teaching Hospital Braunschweig, Braunschweig, Germany. Hauptthemen laut MEDLINE: Hemodiafiltration, Dialysis Solutions, Kidney Failure, Chronic. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 12 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

High‐dose online HDF compared with hemodialysis reduces all‐cause mortality, but increases water consumption. We postulate that the conundrum could be solved by individualizing dialysate flow.

Methods

Thirty CKD5D patients undergoing online HDF post‐dilution were changed from fixed dialysate flow (500 mL/min) to individualized dialysate flow, resulting in a dialysate flow that was always 20% higher than the blood flow. Kt / V urea , K + , Phos, and pill burden in one quater pre as well as four quarters after the change were evaluated.

Results

The practice change resulted in a reduction of the fixed dialysate flow from 500 to 360 (240–383) mL/min. Parameters of dialysis adequacy did not change; neither did the medication.

Conclusion

An individualization of the dialysate flow saved about 36 L dialysate per session; hence, even in light of 23 L used for online HDF, the net burden on the environment could be reduced. Neither dialysis adequacy nor pill burden was negatively affected.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemodiafiltrationDialysis SolutionsKidney Failure, ChronicHumansMaleFemaleMiddle AgedAgedRenal Dialysis

Substanzen

Dialysis Solutions

Autor:innen und Institutionen

  • Medical Clinic V, Nephrology, Rheumatology, and Blood Purification, Academic Teaching Hospital Braunschweig, Braunschweig, Germany.

ORCID-Profile: Kijanosh Lehmann, Jan T Kielstein

Zitieren

APA

David, V., Au, B., Meyer, T. M., Lehmann, K., & Kielstein, J. T. (2025). Smart, Individualized Green Hemodiafiltration ‐ Effects of Individualized Dialysate Flow Through AutoFlow on Dialysate Consumption and Parameters of Dialysis Adequacy in the Setting of Online Hemodiafiltration. Therapeutic Apheresis and Dialysis. https://doi.org/10.1111/1744-9987.70072

Vancouver

David V, Au B, Meyer TM, Lehmann K, Kielstein JT. Smart, Individualized Green Hemodiafiltration ‐ Effects of Individualized Dialysate Flow Through AutoFlow on Dialysate Consumption and Parameters of Dialysis Adequacy in the Setting of Online Hemodiafiltration. Therapeutic Apheresis and Dialysis. 2025. doi:10.1111/1744-9987.70072

BibTeX
@article{David2025,
  author = {David, Vanessa and Au, Bastian and Meyer, Torsten M and Lehmann, Kijanosh and Kielstein, Jan T},
  title = {Smart, Individualized Green Hemodiafiltration ‐ Effects of Individualized Dialysate Flow Through AutoFlow on Dialysate Consumption and Parameters of Dialysis Adequacy in the Setting of Online Hemodiafiltration},
  journal = {Therapeutic Apheresis and Dialysis},
  year = {2025},
  doi = {10.1111/1744-9987.70072}
}
Kostenlos starten →