Performance of X-Ray Photon-Counting Scintillation Detectors Under Pile-Up Conditions at 60 keV

Stefan J. van der Sar, Dennis R. Schaart

IEEE Transactions on Radiation and Plasma Medical Sciences · 2025-07 · 9 (6) · 708-720

DOI: 10.1109/trpms.2025.3532592 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Radiation and Plasma Medical Sciences, 2025-07. Verfasst von Stefan J. van der Sar und 1 weiteren Autor:innen. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We investigate silicon photomultiplier (SiPM)-based scintillation detectors for medical X-ray photon-counting applications, where the input count rate (ICR) can reach a few Mcps/mm2 in cone-beam CT for radiotherapy, for example, up to a few hundred Mcps/mm2 in diagnostic CT. Thus, pulse pile-up can severely distort the measurement of counts and energies. Here, we experimentally evaluate the counting and spectral performance of SiPM-based scintillation detectors at 60 keV as a function of ICR/pile-up level. We coupled$0.9\times 0.9\times 3.5~{\mathrm { mm}}^{3}$LYSO:Ce and$0.9\times 0.9\times 4.5~{\mathrm { mm}}^{3}$YAP:Ce scintillators to$1.0\times 1.0~{\mathrm { mm}}^{2}$ultrafast SiPMs and exposed these single-pixel detectors to a 10-GBq Am-241 source. We varied ICR from 0 to 5 Mcps/pixel and studied detector performance for paralyzable-like (p-like) and nonparalyzable-like (np-like) counting algorithms, after applying a second-order low-pass filter with cut-off frequencies$f_{\mathrm { c}}$of 5, 10, or 20 MHz to the pulse trains. Counting performance was quantified by the output count rate (OCR) and the count-rate loss factor (CRLF). In addition to the traditional spectral performance measure of the full-width-at-half-maximum (FWHM) energy resolution at low ICR, we propose the spectral degradation factor (SDF) to quantify spectral effects of pile-up at any ICR. Best counting performance is obtained with np-like counting and$f_{\mathrm { c}}{=}$20 MHz, for which the count-rate loss is at most 10% in the investigated range of ICRs, whereas p-like counting yields best spectral performance. Due to less pile-up, the fastest pulses obtained with$f_{\mathrm { c}}{=}$20 MHz already provide the best SDF values at ICRs of a few Mcps/pixel, despite their worse low-rate energy resolution. Hence, spectral performance under pile-up conditions appears to benefit more from substantially faster pulses than a somewhat better low-rate energy resolution. Moreover, we show that the pulse shape of SiPM-based detectors allows to improve spectral performance under pile-up conditions using dedicated peak detection windows.

Zitieren

APA

Sar, S. J. V. D., & Schaart, D. R. (2025). Performance of X-Ray Photon-Counting Scintillation Detectors Under Pile-Up Conditions at 60 keV. IEEE Transactions on Radiation and Plasma Medical Sciences, 9(6), 708-720. https://doi.org/10.1109/trpms.2025.3532592

Vancouver

Sar SJVD, Schaart DR. Performance of X-Ray Photon-Counting Scintillation Detectors Under Pile-Up Conditions at 60 keV. IEEE Transactions on Radiation and Plasma Medical Sciences. 2025;9(6):708-720. doi:10.1109/trpms.2025.3532592

BibTeX
@article{Sar2025,
  author = {Sar, Stefan J. van der and Schaart, Dennis R.},
  title = {Performance of X-Ray Photon-Counting Scintillation Detectors Under Pile-Up Conditions at 60 keV},
  journal = {IEEE Transactions on Radiation and Plasma Medical Sciences},
  year = {2025},
  volume = {9},
  number = {6},
  pages = {708--720},
  doi = {10.1109/trpms.2025.3532592}
}
Kostenlos starten →