Performance Study and Circuit Simulation for R12699-406-M4 Photomultiplier Tube Base
IEEE Transactions on Nuclear Science · 2026-06 · 73 (6) · 2511-2519
DOI: 10.1109/tns.2026.3689710 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Transactions on Nuclear Science, 2026-06. Verfasst von Peiyuan Chen und 12 weiteren Autor:innen. Die Arbeit nennt 29 Literaturangaben. Gefördert von: National Key Research and Development Program of China, National Natural Science Foundation of China, Office of Science and Technology, Shanghai Municipal Government.
Abstract (Originalsprache)
The next-generation liquid xenon (LXe) experiments like PandaX-xT target an energy range from sub-keV to multi-MeV to address the requirement of multiple physics searches. The Hamamatsu R12699-406-M4 photomultiplier tubes (PMTs) were developed and selected as photon sensors for PandaX-xT. Their voltage-divider base is optimized for a broad dynamic range, from single-photoelectron (SPE) sensitivity to 30 nC collected charge, corresponding to the 2.458 MeV $Q$ value of 136Xe neutrinoless double beta decay (NLDBD). Using a dedicated test bench, we characterize the saturation and suppression responses of R12699-406-M4 PMTs with this base design. Based on measured PMT-base responses, we develop a circuit simulation model that accurately reproduces the physical mechanisms underlying these effects with key parameters tuned via experimental data. The combined simulation and bench-test approach guides base design and optimization, enabling improved detector dynamic range and supporting future saturation and suppression correction studies in data analysis.
Förderung
- National Key Research and Development Program of China
- National Natural Science Foundation of China
- Office of Science and Technology, Shanghai Municipal Government
- Natural Science Foundation of Shanghai
- Innovation Funds of LingChuang Research Project of China National Nuclear Corporation (CNCC)
- Fundamental Research Funds for the Central Universities
- Chinese Academy of Sciences Center for Excellence in Particle Physics
- Thomas and Linda Lau Family Foundation
Zitieren
APA
Chen, P., Huang, H., Han, K., Liu, J., Liu, Y., Wang, G., Wang, S., Wu, W., Yan, B., Ye, P., Yu, H., Zhou, J., & Zhou, Z. (2026). Performance Study and Circuit Simulation for R12699-406-M4 Photomultiplier Tube Base. IEEE Transactions on Nuclear Science, 73(6), 2511-2519. https://doi.org/10.1109/tns.2026.3689710
Vancouver
Chen P, Huang H, Han K, Liu J, Liu Y, Wang G, et al. Performance Study and Circuit Simulation for R12699-406-M4 Photomultiplier Tube Base. IEEE Transactions on Nuclear Science. 2026;73(6):2511-2519. doi:10.1109/tns.2026.3689710
BibTeX
@article{Chen2026,
author = {Chen, Peiyuan and Huang, Houqi and Han, Ke and Liu, Jianglai and Liu, Yang and Wang, Guanbo and Wang, Shaobo and Wu, Weihao and Yan, Binbin and Ye, Peihua and Yu, Haosong and Zhou, Jiaxu and Zhou, Zhizhen},
title = {Performance Study and Circuit Simulation for R12699-406-M4 Photomultiplier Tube Base},
journal = {IEEE Transactions on Nuclear Science},
year = {2026},
volume = {73},
number = {6},
pages = {2511--2519},
doi = {10.1109/tns.2026.3689710}
}
EUVIMED