EUVIMEDEUVIMED entdecken
Crossref

Image-Based Backlash Width Identification for Eye-in-Hand Robotized Flexible Endoscopes

Thibault Poignonec, Philippe Zanne, Nabil Zemiti, Bernard Bayle, Florent Nageotte

IEEE Transactions on Medical Robotics and Bionics · 2026-08 · 8 (3) · 1122-1132

DOI: 10.1109/tmrb.2026.3712813 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Medical Robotics and Bionics, 2026-08. Verfasst von Thibault Poignonec und 4 weiteren Autor:innen. Die Arbeit nennt 27 Literaturangaben. Gefördert von: French State Funds managed by the Agence Nationale de la RechercheLabex (ANR) within France 2030 Program for the Labex CAMI, Institut Thématique Interdisciplinaire (ITI) HealthTech, TIRREX.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Robotic flexible endoscopes allow for more dexterous and ergonomic control than their manual counterparts. However, they are usually actuated by tendon-sheath mechanisms that exhibit backlash-like hysteretic behavior when operated. Various methods have been proposed to perform the model identification of these robots for control purposes, including the compensation of mechanical backlash, but they usually consider that the robotic system is visible to an endoscopic camera, which is not the case in robotic flexible endoscopes. In this paper, we introduce a novel method for automatic in-situ backlash identification in an eye-in-hand configuration, relying on embedded image measurements only. We propose to track scene landmarks in the image in order to estimate simultaneously their 3D positions and the parameters of the backlash model. The resulting learning problem is solved using a discontinuous extended Kalman filter to cope with the non-smooth model. The proposed approach is studied through simulations and then validated experimentally on a robotized endoscopic platform. The results show that the method allows to estimate non-constant backlash using only the standard embedded endoscopic camera and is thus applicable on-site for most flexible endoscopic robotic systems.

Förderung

  • French State Funds managed by the Agence Nationale de la RechercheLabex (ANR) within France 2030 Program for the Labex CAMI
  • Institut Thématique Interdisciplinaire (ITI) HealthTech
  • TIRREX

Zitieren

APA

Poignonec, T., Zanne, P., Zemiti, N., Bayle, B., & Nageotte, F. (2026). Image-Based Backlash Width Identification for Eye-in-Hand Robotized Flexible Endoscopes. IEEE Transactions on Medical Robotics and Bionics, 8(3), 1122-1132. https://doi.org/10.1109/tmrb.2026.3712813

Vancouver

Poignonec T, Zanne P, Zemiti N, Bayle B, Nageotte F. Image-Based Backlash Width Identification for Eye-in-Hand Robotized Flexible Endoscopes. IEEE Transactions on Medical Robotics and Bionics. 2026;8(3):1122-1132. doi:10.1109/tmrb.2026.3712813

BibTeX
@article{Poignonec2026,
  author = {Poignonec, Thibault and Zanne, Philippe and Zemiti, Nabil and Bayle, Bernard and Nageotte, Florent},
  title = {Image-Based Backlash Width Identification for Eye-in-Hand Robotized Flexible Endoscopes},
  journal = {IEEE Transactions on Medical Robotics and Bionics},
  year = {2026},
  volume = {8},
  number = {3},
  pages = {1122--1132},
  doi = {10.1109/tmrb.2026.3712813}
}
Kostenlos starten →