EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

An Effectiveness Study Across Baseline and Learning-Based Force Estimation Methods on the da Vinci Research Kit Si System

Hao Yang, Ayberk Acar, Keshuai Xu, Anton Deguet, Peter Kazanzides, Jie Ying Wu

IEEE Transactions on Medical Robotics and Bionics · 2025-11 · 7 (4) · 1393-1397

DOI: 10.1109/tmrb.2025.3589744 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Medical Robotics and Bionics, 2025-11. Verfasst von Hao Yang und 5 weiteren Autor:innen. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Intuitive Surgical Technology Research Grant.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Robot-assisted minimally invasive surgery, such as through the da Vinci systems, improves precision and patient outcomes. However, da Vinci systems prior to da Vinci 5 lacked direct force-sensing capabilities, forcing surgeons to operate without the haptic feedback they get through laparoscopy. Our prior work restored force sensing through machine learning-based force estimation for an open-source surgical robotics research platform, the da Vinci Research Kit (dVRK) Classic. This study extends our previous method to the newer dVRK system, the dVRK-Si. Additionally, we benchmark the performance of the learning-based algorithm against baseline methods (which make simplifying assumptions on the torque) to study how the two systems differ. In both systems, the learning-based method outperforms baselines, but the difference is much larger in the dVRK-Si. Nonetheless, dVRK-Si force estimation accuracy lags behind the dVRK Classic, with Root Mean Square Error (RMSE) 2 to 3 times higher. Further analysis reveals suboptimal PID control in the dVRK-Si. We hypothesize that this is because, unlike the dVRK Classic, the dVRK-Si is not mechanically balanced and exhibits more complex internal dynamics. This study advances the understanding of learning-based force estimation and is the first work to implement learning-based dynamics estimation of the new dVRK-Si system.

Förderung

  • Intuitive Surgical Technology Research Grant

Zitieren

APA

Yang, H., Acar, A., Xu, K., Deguet, A., Kazanzides, P., & Wu, J. Y. (2025). An Effectiveness Study Across Baseline and Learning-Based Force Estimation Methods on the da Vinci Research Kit Si System. IEEE Transactions on Medical Robotics and Bionics, 7(4), 1393-1397. https://doi.org/10.1109/tmrb.2025.3589744

Vancouver

Yang H, Acar A, Xu K, Deguet A, Kazanzides P, Wu JY. An Effectiveness Study Across Baseline and Learning-Based Force Estimation Methods on the da Vinci Research Kit Si System. IEEE Transactions on Medical Robotics and Bionics. 2025;7(4):1393-1397. doi:10.1109/tmrb.2025.3589744

BibTeX
@article{Yang2025,
  author = {Yang, Hao and Acar, Ayberk and Xu, Keshuai and Deguet, Anton and Kazanzides, Peter and Wu, Jie Ying},
  title = {An Effectiveness Study Across Baseline and Learning-Based Force Estimation Methods on the da Vinci Research Kit Si System},
  journal = {IEEE Transactions on Medical Robotics and Bionics},
  year = {2025},
  volume = {7},
  number = {4},
  pages = {1393--1397},
  doi = {10.1109/tmrb.2025.3589744}
}
Kostenlos starten →