EUVIMEDEUVIMED entdecken
Open Access

Steady-Hand Eye Robot 3.0: Optimization and Benchtop Evaluation for Subretinal Injection

Alireza Alamdar, David E Usevitch, Jiahao Wu, Russell H Taylor, Peter Gehlbach, Iulian Iordachita

IEEE Transactions on Medical Robotics and Bionics · 2024-02 · 6 (1) · 135-145

DOI: 10.1109/tmrb.2023.3336975 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Medical Robotics and Bionics, 2024-02. Verfasst von Alireza Alamdar und 5 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory for Computational Sensing and Robotics (LCSR), Johns Hopkins University, Baltimore, MD USA. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, U.S. National Institutes of Health, Johns Hopkins University Internal Funds.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Subretinal injection methods and other procedures for treating retinal conditions and diseases (many considered incurable) have been limited in scope due to limited human motor control. This study demonstrates the next generation, cooperatively controlled Steady-Hand Eye Robot (SHER 3.0), a precise and intuitive-to-use robotic platform achieving clinical standards for targeting accuracy and resolution for subretinal injections. The system design and basic kinematics are reported and a deflection model for the incorporated delta stage and validation experiments are presented. This model optimizes the delta stage parameters, maximizing the global conditioning index and minimizing torsional compliance. Five tests measuring accuracy, repeatability, and deflection show the optimized stage design achieves a tip accuracy of < 30 μm, tip repeatability of 9.3 μm and 0.02°, and deflections between 20-350 μm/N. Future work will use updated control models to refine tip positioning outcomes and will be tested on in vivo animal models.

Schlagwörter

Stichwörter der Autor:innen

ophthalmologyoptimizationsubretinal injectionsurgical roboticssystem design and analysis

Autor:innen und Institutionen

  • Laboratory for Computational Sensing and Robotics (LCSR), Johns Hopkins University, Baltimore, MD USA.
  • Wilmer Eye Institute, Johns Hopkins University, Baltimore, MD 21287, USA.

Förderung

  • NIBIB NIH HHS
  • U.S. National Institutes of Health
  • Johns Hopkins University Internal Funds

Zitieren

APA

Alamdar, A., Usevitch, D. E., Wu, J., Taylor, R. H., Gehlbach, P., & Iordachita, I. (2024). Steady-Hand Eye Robot 3.0: Optimization and Benchtop Evaluation for Subretinal Injection. IEEE Transactions on Medical Robotics and Bionics, 6(1), 135-145. https://doi.org/10.1109/tmrb.2023.3336975

Vancouver

Alamdar A, Usevitch DE, Wu J, Taylor RH, Gehlbach P, Iordachita I. Steady-Hand Eye Robot 3.0: Optimization and Benchtop Evaluation for Subretinal Injection. IEEE Transactions on Medical Robotics and Bionics. 2024;6(1):135-145. doi:10.1109/tmrb.2023.3336975

BibTeX
@article{Alamdar2024,
  author = {Alamdar, Alireza and Usevitch, David E and Wu, Jiahao and Taylor, Russell H and Gehlbach, Peter and Iordachita, Iulian},
  title = {Steady-Hand Eye Robot 3.0: Optimization and Benchtop Evaluation for Subretinal Injection},
  journal = {IEEE Transactions on Medical Robotics and Bionics},
  year = {2024},
  volume = {6},
  number = {1},
  pages = {135--145},
  doi = {10.1109/tmrb.2023.3336975}
}
Kostenlos starten →