Design and Validation of a Novel Robotic Neck Brace for Cervical Traction
IEEE/ASME Transactions on Mechatronics · 2024-08 · 29 (4) · 3092-3099
DOI: 10.1109/tmech.2024.3402614 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE/ASME Transactions on Mechatronics, 2024-08. Verfasst von Priya Kulkarni und 1 weiteren Autor:innen. Beteiligte Einrichtung: P. Kulkarni and S. K. Agrawal are with the Robotics and Rehabilitation Laboratory, Department of Mechanical Engineering, Columbia University, New Yor…. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NICHD NIH HHS, ALS Association, Columbia University School of Engineering seed funding.
Abstract (Originalsprache)
Cervical traction is a common and effective treatment for degenerative disk diseases and pain in the cervical spine. However, the manual or mechanical methods of applying traction to the head-neck are limited due to variability in the applied forces and orientation of the head-neck relative to the shoulder during the procedure. Current robotic neck braces are not designed to provide independent rotation angles and independent vertical translation, or traction, to the brace end-effector connected to the head, making them unsuitable for traction application. This work proposes a novel architecture of a robotic neck brace, which can provide vertical traction to the head while keeping the head in a prescribed orientation, with flexion and lateral bending angles. In this paper, the kinematics of the end-effector attached to the head relative to a coordinate frame on the shoulders are described as well as the velocity kinematics and force control. The paper also describes benchtop experiments designed to validate the position control and the ability of the brace to provide a vertical traction force. It was shown that the maximum achievable end-effector orientations are 16° in flexion, 13.9° in extension, and ± 6.5° in lateral bending. The kinematic model of the active brace was validated using an independent motion capture system with a maximum root mean square error of 2.41°. In three different orientations of the end-effector, neutral, flexed, and laterally bent, the brace was able to provide a consistent upward traction force during intermittent force application. In these configurations, the force error has standard deviations of 0.55, 0.29, and 0.07N, respectively. This work validates the mechanism's ability to achieve a range of head orientations and provide consistent upward traction force in these orientations, making it a promising intervention tool in cases of cervical disk degeneration.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- P. Kulkarni and S. K. Agrawal are with the Robotics and Rehabilitation Laboratory, Department of Mechanical Engineering, Columbia University, New York, NY 10027 USA.
- S. K. Agrawal is with the Department of Rehabilitation and Regenerative Medicine, Columbia University Medical Center, New York, NY 10032 USA.
Förderung
- NICHD NIH HHS
- ALS Association
- Columbia University School of Engineering seed funding
- New York State Spinal Cord Injury Research Board
- NYS SCIRB Predoctoral Fellowship
- National Institutes of Health
Zitieren
APA
Kulkarni, P., & Agrawal, S. K. (2024). Design and Validation of a Novel Robotic Neck Brace for Cervical Traction. IEEE/ASME Transactions on Mechatronics, 29(4), 3092-3099. https://doi.org/10.1109/tmech.2024.3402614
Vancouver
Kulkarni P, Agrawal SK. Design and Validation of a Novel Robotic Neck Brace for Cervical Traction. IEEE/ASME Transactions on Mechatronics. 2024;29(4):3092-3099. doi:10.1109/tmech.2024.3402614
BibTeX
@article{Kulkarni2024,
author = {Kulkarni, Priya and Agrawal, Sunil K},
title = {Design and Validation of a Novel Robotic Neck Brace for Cervical Traction},
journal = {IEEE/ASME Transactions on Mechatronics},
year = {2024},
volume = {29},
number = {4},
pages = {3092--3099},
doi = {10.1109/tmech.2024.3402614}
}
EUVIMED