Synchronous Dynamic Vibration Calibration for Sensitivity Matrix of Triaxial Accelerometers Based on Coupling Measurement Error Correction
IEEE Transactions on Instrumentation and Measurement · 2026 · 75 · 1006415-1006415
DOI: 10.1109/tim.2026.3712890 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Transactions on Instrumentation and Measurement, 2026. Verfasst von Xufei Zhang und 4 weiteren Autor:innen. Die Arbeit nennt 47 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Basic Research Program of Shanxi Province.
Abstract (Originalsprache)
Triaxial accelerometers are widely used in engineering applications requiring high-precision multiaxis vibration measurements, and their measurement accuracy strongly depends on reliable calibration. Traditional single-axis calibration methods cannot characterize the interaxis coupling effects of triaxial accelerometers. Meanwhile, triaxial synchronous dynamic calibration systems based on laser interferometry are costly and complex, thereby limiting practical implementation. Therefore, developing a low-cost, easy-to-operate, and high-precision calibration method for triaxial accelerometers is of significant research and engineering value. In this study, a novel synchronous dynamic calibration system based on a coupling measurement error correction method is proposed to achieve high-precision calibration of the triaxial accelerometer sensitivity matrix. First, the coupling measurement errors introduced by three low-cost reference single-axis accelerometers are analyzed in triaxial vibration excitation measurement. Then, a correction method is proposed to reconstruct accurate triaxial vibration acceleration signals. Based on the correction method, a synchronous dynamic calibration system for the sensitivity matrix of the triaxial accelerometer is established. Finally, the correctness and effectiveness of the proposed correction method and synchronous dynamic calibration system are verified through experimental tests. The results indicate that the relative errors between the experimentally calibrated axial sensitivities and the corresponding factory-calibrated values range from 0.08 % to 3.09 % at test frequencies of 0.8, 1, and 2 Hz. The combined and expanded calibration uncertainties are estimated to be 1.91 % and 3.83 %, respectively. Further validation experiments show that the validation errors obtained using the calibrated sensitivity matrix do not exceed 4.60 %. The proposed method and system provide a feasible technical solution for cost-effective and high-precision calibration of triaxial accelerometer sensitivity matrices.
Förderung
- National Natural Science Foundation of China
- Basic Research Program of Shanxi Province
Zitieren
APA
Zhang, X., Xu, T., Cao, C., Ma, J., & Sun, S. (2026). Synchronous Dynamic Vibration Calibration for Sensitivity Matrix of Triaxial Accelerometers Based on Coupling Measurement Error Correction. IEEE Transactions on Instrumentation and Measurement, 75, 1006415-1006415. https://doi.org/10.1109/tim.2026.3712890
Vancouver
Zhang X, Xu T, Cao C, Ma J, Sun S. Synchronous Dynamic Vibration Calibration for Sensitivity Matrix of Triaxial Accelerometers Based on Coupling Measurement Error Correction. IEEE Transactions on Instrumentation and Measurement. 2026;75:1006415-1006415. doi:10.1109/tim.2026.3712890
BibTeX
@article{Zhang2026,
author = {Zhang, Xufei and Xu, Tuo and Cao, Chengqian and Ma, Jie and Sun, Shengli},
title = {Synchronous Dynamic Vibration Calibration for Sensitivity Matrix of Triaxial Accelerometers Based on Coupling Measurement Error Correction},
journal = {IEEE Transactions on Instrumentation and Measurement},
year = {2026},
volume = {75},
pages = {1006415--1006415},
doi = {10.1109/tim.2026.3712890}
}