A Handheld Wireless Force-Sensing-Ultrasound Acquisition System for Measuring Real-Time Contact Compressive Forces

Hoang B. V. Dinh, Lawrence H. Le, Camila E. Orsso, Vu-Duc Nguyen, Carla M. Prado, Edmond Lou

IEEE Transactions on Instrumentation and Measurement · 2026 · 75 · 4008110-4008110

DOI: 10.1109/tim.2026.3693736 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Instrumentation and Measurement, 2026. Verfasst von Hoang B. V. Dinh und 5 weiteren Autor:innen. Die Arbeit nennt 32 Literaturangaben. Gefördert von: Alberta Innovates and in part by the Natural Sciences and Engineering Research Council of Canada (NSERC) under the NSERC Alliance—Alberta Innovates Advanced Pr….

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Ultrasound (US) is an emerging, bedside-compatible imaging modality for assessing skeletal muscle mass and quality. However, image quality and soft-tissue compression during scanning are highly operator-dependent, thereby affecting measurement accuracy and reproducibility. To address this limitation, a wireless force-sensing system was developed and integrated into a portable US probe to measure the applied interface force. By measuring probe force, the system enables characterization of the relationship between applied load and soft-tissue compression, providing a basis for distinguishing externally induced deformation from true muscle changes. This study reports the development of a low-cost, miniature, and wireless force-sensing module and its integration into a handheld US probe. Three housing designs were evaluated to assess their effects on sensor sensitivity, and an optimized soft, flexible housing with a stiff bead pad was selected. The final system achieved a sensitivity of 364 mV/N, a repeatability error of 0.383%, and a reproducibility error of 0.436%. Formal calibration demonstrated worst case combined relative standard uncertainties of 0.085 and 0.113 for the two sensors, corresponding to expanded uncertainties (k = 2) of 0.171 and 0.226, respectively, over an operating force range of 0.294–2.45 N. An in vivo pilot study in a healthy volunteer further showed a strong correlation (R ${}^{2} =0.91$ ) between applied force and muscle thickness deformation, demonstrating the feasibility of standardized muscle US measurements.

Förderung

  • Alberta Innovates and in part by the Natural Sciences and Engineering Research Council of Canada (NSERC) under the NSERC Alliance—Alberta Innovates Advanced Pr…

Zitieren

APA

Dinh, H. B. V., Le, L. H., Orsso, C. E., Nguyen, V., Prado, C. M., & Lou, E. (2026). A Handheld Wireless Force-Sensing-Ultrasound Acquisition System for Measuring Real-Time Contact Compressive Forces. IEEE Transactions on Instrumentation and Measurement, 75, 4008110-4008110. https://doi.org/10.1109/tim.2026.3693736

Vancouver

Dinh HBV, Le LH, Orsso CE, Nguyen V, Prado CM, Lou E. A Handheld Wireless Force-Sensing-Ultrasound Acquisition System for Measuring Real-Time Contact Compressive Forces. IEEE Transactions on Instrumentation and Measurement. 2026;75:4008110-4008110. doi:10.1109/tim.2026.3693736

BibTeX
@article{Dinh2026,
  author = {Dinh, Hoang B. V. and Le, Lawrence H. and Orsso, Camila E. and Nguyen, Vu-Duc and Prado, Carla M. and Lou, Edmond},
  title = {A Handheld Wireless Force-Sensing-Ultrasound Acquisition System for Measuring Real-Time Contact Compressive Forces},
  journal = {IEEE Transactions on Instrumentation and Measurement},
  year = {2026},
  volume = {75},
  pages = {4008110--4008110},
  doi = {10.1109/tim.2026.3693736}
}
Kostenlos starten →