EUVIMEDEUVIMED entdecken
Crossref

Integrated Pneumatic Servo Module for Mechanical Ventilation With ADRC and RLS-Based Feedforward Control

Xiaochuan Ming, Shuai Ren, Xiaoyi Wang, Tao Wang, Yuying Lv, Zujin Luo

IEEE Transactions on Industrial Electronics · 2026-09 · 73 (9) · 12691-12701

DOI: 10.1109/tie.2026.3675040 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Industrial Electronics, 2026-09. Verfasst von Xiaochuan Ming und 5 weiteren Autor:innen. Die Arbeit nennt 37 Literaturangaben. Gefördert von: Beijing Natural Science Foundation, National Natural Science Foundation of China, National Science and Technology Major Project.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This article presents a miniaturized integrated pneumatic servo module with hybrid adaptive control for mechanical ventilation. Unlike conventional ventilators requiring three or more actuators, the proposed design achieves air-oxygen blending and pressure/flow regulation using only two proportional solenoid valves. The control architecture adopts dual closed-loop control, with parallel active disturbance rejection control for the inner flow loops and PI control for the outer pressure loop. Online recursive least squares identification of respiratory mechanics deployed for adaptive feedforward compensation. Fuzzy control was adopted for positive end-expiratory pressure (PEEP) regulation. Experimental results demonstrate a 300 ms rise time with zero overshoot in pressure tracking, FiO ${}_{\mathbf{2}}$ control accuracy within 5% across the 30–90% oxygen concentration range, PEEP control accuracy within $\boldsymbol{\pm}$ 1.5%, and robust performance under varying respiratory mechanics. The integrated design reduces system complexity while maintaining clinical-grade performance, offering a cost-effective solution for resource-constrained healthcare settings.

Förderung

  • Beijing Natural Science Foundation
  • National Natural Science Foundation of China
  • National Science and Technology Major Project

Zitieren

APA

Ming, X., Ren, S., Wang, X., Wang, T., Lv, Y., & Luo, Z. (2026). Integrated Pneumatic Servo Module for Mechanical Ventilation With ADRC and RLS-Based Feedforward Control. IEEE Transactions on Industrial Electronics, 73(9), 12691-12701. https://doi.org/10.1109/tie.2026.3675040

Vancouver

Ming X, Ren S, Wang X, Wang T, Lv Y, Luo Z. Integrated Pneumatic Servo Module for Mechanical Ventilation With ADRC and RLS-Based Feedforward Control. IEEE Transactions on Industrial Electronics. 2026;73(9):12691-12701. doi:10.1109/tie.2026.3675040

BibTeX
@article{Ming2026,
  author = {Ming, Xiaochuan and Ren, Shuai and Wang, Xiaoyi and Wang, Tao and Lv, Yuying and Luo, Zujin},
  title = {Integrated Pneumatic Servo Module for Mechanical Ventilation With ADRC and RLS-Based Feedforward Control},
  journal = {IEEE Transactions on Industrial Electronics},
  year = {2026},
  volume = {73},
  number = {9},
  pages = {12691--12701},
  doi = {10.1109/tie.2026.3675040}
}
Kostenlos starten →