A Fully-Integrated 0.068-mm 3 Implantable Pressure Sensing Device With Wireless Energy Harvesting and Data Telemetry
IEEE Transactions on Biomedical Circuits and Systems · 2026
DOI: 10.1109/tbcas.2025.3586009 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Transactions on Biomedical Circuits and Systems, 2026. Verfasst von Zehua Lan und 5 weiteren Autor:innen. Hauptthemen laut MEDLINE: Telemetry, Wireless Technology, Prostheses and Implants. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 32 Literaturangaben. Gefördert von: National Key Research and Development Program of China, Beijing Engineering Research Center.
Abstract (Originalsprache)
This paper reports a fully-integrated sub-0.1 mm3 wireless pressure sensing device for implantable applications. The miniature device integrates a customized system-on-a-chip (SoC) and an off-the-shelf half-bridge piezoresistive pressure transducer, eliminating off-chip passive components. The SoC mainly comprises a resistance-to-time converter, a 915 MHz inductively coupled energy harvester with an on-chip coil, and a backscatter telemetry. Key innovations enabling low power, small size and high precision include: (1) A source-input common-gate amplifier based R-V converter, that reuses the transducer's bias current, (2) Advanced noise management via chopper stabilization and supply noise cancellation, and (3) A compact high-Q on-chip multi-layer stacked coil design for wireless link. The active circuits consume 9.75 $\boldsymbol{\mu}$W, fully supplied by the energy harvested wirelessly through the on-chip coil. The sensing data is transmitted wirelessly to an external recorder through the RF backscatter link. Fabricated in a 65-nm CMOS technology, the SoC occupies a die area of 400 µm × 490 µm, and the entire fully-integrated sensor has a volume of only 0.068 mm3, enabling syringe injection through a ≤0.5 mm needle. Experiments with the sensing device covered by pork have demonstrated that the device can operate at an implant depth of up to 10 mm with excellent misalignment tolerance. It offers a pressure sensing resolution of 3.1 mmHg over a relative pressure range of 0-200 mmHg and a temperature sensing resolution of 0.18°C.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Förderung
- National Key Research and Development Program of China
- Beijing Engineering Research Center
Zitieren
APA
Lan, Z., Shi, J., Hao, J., Wang, Z., Guo, Y., & Jiang, H. (2026). A Fully-Integrated 0.068-mm 3 Implantable Pressure Sensing Device With Wireless Energy Harvesting and Data Telemetry. IEEE Transactions on Biomedical Circuits and Systems. https://doi.org/10.1109/tbcas.2025.3586009
Vancouver
Lan Z, Shi J, Hao J, Wang Z, Guo Y, Jiang H. A Fully-Integrated 0.068-mm 3 Implantable Pressure Sensing Device With Wireless Energy Harvesting and Data Telemetry. IEEE Transactions on Biomedical Circuits and Systems. 2026. doi:10.1109/tbcas.2025.3586009
BibTeX
@article{Lan2026,
author = {Lan, Zehua and Shi, Jiahua and Hao, Jiayue and Wang, Zhihua and Guo, Yanshu and Jiang, Hanjun},
title = {A Fully-Integrated 0.068-mm 3 Implantable Pressure Sensing Device With Wireless Energy Harvesting and Data Telemetry},
journal = {IEEE Transactions on Biomedical Circuits and Systems},
year = {2026},
doi = {10.1109/tbcas.2025.3586009}
}