32 Element Piezoelectric Micromachined Ultrasound Transducer (PMUT) Phased Array for Neuromodulation
IEEE Open Journal of Ultrasonics, Ferroelectrics, and Frequency Control · 2022
DOI: 10.1109/ojuffc.2022.3196823 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Open Journal of Ultrasonics, Ferroelectrics, and Frequency Control, 2022. Verfasst von Pannawit Tipsawat und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Materials Science and Engineering and Materials Research Institute, The Pennsylvania State University, University Park, PA 16802 USA. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NEI NIH HHS, National Institutes of Health, Development and Promotion of Science and Technology Talents Project (DPST) Scholarship, Thailand.
Abstract (Originalsprache)
Interest in utilizing ultrasound (US) transducers for non-invasive neuromodulation treatment, including for low intensity transcranial focused ultrasound stimulation (tFUS), has grown rapidly. The most widely demonstrated US transducers for tFUS are either bulk piezoelectric transducers or capacitive micromachine transducers (CMUT) which require high voltage excitation to operate. In order to advance the development of the US transducers towards small, portable devices for safe tFUS at large scale, a low voltage array of US transducers with beam focusing and steering capability is of interest. This work presents the design methodology, fabrication, and characterization of 32-element phased array piezoelectric micromachined ultrasound transducers (PMUT) using 1.5 μm thick Pb(Zr0.52 Ti0.48)O3 films doped with 2 mol% Nb. The electrode/piezoelectric/electrode stack was deposited on a silicon on insulator (SOI) wafer with a 2 μm silicon device layer that serves as the passive elastic layer for bending-mode vibration. The fabricated 32-element PMUT has a central frequency at 1.4 MHz. Ultrasound beam focusing and steering (through beamforming) was demonstrated where the array was driven with 14.6 V square unipolar pulses. The PMUT generated a maximum peak-to-peak focused acoustic pressure output of 0.44 MPa at a focal distance of 20 mm with a 9.2 mm and 1 mm axial and lateral resolution, respectively. The maximum pressure is equivalent to a spatial-peak pulse-average intensity of 1.29 W/cm2, which is suitable for tFUS application.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Materials Science and Engineering and Materials Research Institute, The Pennsylvania State University, University Park, PA 16802 USA.
- Department of Electrical Engineering, The Pennsylvania State University, University Park, PA 16802 USA.
ORCID-Profile: Pannawit Tipsawat, Sheikh Jawad Ilham, Zeinab Kashani, Mehdi Kiani, Susan Trolier-McKinstry
Förderung
- NEI NIH HHS
- National Institutes of Health
- Development and Promotion of Science and Technology Talents Project (DPST) Scholarship, Thailand
Zitieren
APA
Tipsawat, P., Ilham, S. J., Yang, J. I., Kashani, Z., Kiani, M., & Trolier-McKinstry, S. (2022). 32 Element Piezoelectric Micromachined Ultrasound Transducer (PMUT) Phased Array for Neuromodulation. IEEE Open Journal of Ultrasonics, Ferroelectrics, and Frequency Control. https://doi.org/10.1109/ojuffc.2022.3196823
Vancouver
Tipsawat P, Ilham SJ, Yang JI, Kashani Z, Kiani M, Trolier-McKinstry S. 32 Element Piezoelectric Micromachined Ultrasound Transducer (PMUT) Phased Array for Neuromodulation. IEEE Open Journal of Ultrasonics, Ferroelectrics, and Frequency Control. 2022. doi:10.1109/ojuffc.2022.3196823
BibTeX
@article{Tipsawat2022,
author = {Tipsawat, Pannawit and Ilham, Sheikh Jawad and Yang, Jung In and Kashani, Zeinab and Kiani, Mehdi and Trolier-McKinstry, Susan},
title = {32 Element Piezoelectric Micromachined Ultrasound Transducer (PMUT) Phased Array for Neuromodulation},
journal = {IEEE Open Journal of Ultrasonics, Ferroelectrics, and Frequency Control},
year = {2022},
doi = {10.1109/ojuffc.2022.3196823}
}
EUVIMED