EUVIMEDEUVIMED entdecken
Crossref

A Capillary-Driven Textile Electrochemical Device Enabling Rapid and Continuous Uric Acid Monitoring in Sweat

Toru Nohgi, Huizi Zhang, Qianxi Zhao, Alexander Sakae Kameoka, Sirawit Sornrak, Ling Ding, Hiroshi Kawahira, Jun Kameoka

IEEE Sensors Letters · 2026-09 · 10 (9) · 4503104-4503104

DOI: 10.1109/lsens.2026.3708910 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Sensors Letters, 2026-09. Verfasst von Toru Nohgi und 7 weiteren Autor:innen. Die Arbeit nennt 17 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this study, we developed a textile-based electrochemical sensor for continuous, noninvasive monitoring of uric acid (UA) in sweat. Conventional sweat-sensing platforms often suffer from air bubble formation, flow instability, and reliance on external pumps, limiting continuous sampling and rapid signal refresh. To address these issues, we integrated a fiber-based sweat transport pathway with electrochemical sensing into a single textile platform. By exploiting capillary action within textile fibers, the device enabled continuous and stable sweat transport with a refresh time of 13–15 s. Human exercise experiments demonstrated continuous monitoring of sweat UA and its relationship with blood UA. A positive correlation was observed between sweat UA and blood UA, and cross-correlation analysis revealed a time lag of approximately 3 min (R = 0.8149). In addition, exercise- and recovery-related changes in blood UA were reflected in sweat. The novelty of this work lies in integrating passive capillary-driven sweat transport with electrochemical UA sensing in a textile-based wearable device, thereby overcoming flow instability and pump dependence in conventional sweat-sensing systems.

Zitieren

APA

Nohgi, T., Zhang, H., Zhao, Q., Kameoka, A. S., Sornrak, S., Ding, L., Kawahira, H., & Kameoka, J. (2026). A Capillary-Driven Textile Electrochemical Device Enabling Rapid and Continuous Uric Acid Monitoring in Sweat. IEEE Sensors Letters, 10(9), 4503104-4503104. https://doi.org/10.1109/lsens.2026.3708910

Vancouver

Nohgi T, Zhang H, Zhao Q, Kameoka AS, Sornrak S, Ding L, et al. A Capillary-Driven Textile Electrochemical Device Enabling Rapid and Continuous Uric Acid Monitoring in Sweat. IEEE Sensors Letters. 2026;10(9):4503104-4503104. doi:10.1109/lsens.2026.3708910

BibTeX
@article{Nohgi2026,
  author = {Nohgi, Toru and Zhang, Huizi and Zhao, Qianxi and Kameoka, Alexander Sakae and Sornrak, Sirawit and Ding, Ling and Kawahira, Hiroshi and Kameoka, Jun},
  title = {A Capillary-Driven Textile Electrochemical Device Enabling Rapid and Continuous Uric Acid Monitoring in Sweat},
  journal = {IEEE Sensors Letters},
  year = {2026},
  volume = {10},
  number = {9},
  pages = {4503104--4503104},
  doi = {10.1109/lsens.2026.3708910}
}
Kostenlos starten →