EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Non-Contact Measurement of Blood Oxygen Saturation Using Facial Video Without Reference Values

Soma Sasaki, Norihiro Sugita iD, Takanori Terai, Makoto Yoshizawa iD

IEEE Journal of Translational Engineering in Health and Medicine · 2024 · 12 · 76-83

DOI: 10.1109/jtehm.2023.3318643 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Journal of Translational Engineering in Health and Medicine, 2024. Verfasst von Soma Sasaki und 3 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of EngineeringTohoku University Sendai 9808579 Japan. Hauptthemen laut MEDLINE: Oxygen Saturation, Oximetry. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nakatani Foundation for Advancement of Measuring Technologies in Biomedical Engineering.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The continuous measurement of percutaneous oxygen saturation (SpO2) enables diseases that cause hypoxemia to be detected early and patients' conditions to be monitored. Currently, SpO2 is mainly measured using a pulse oximeter, which, owing to its simplicity, can be used in clinical settings and at home. However, the pulse oximeter requires a sensor to be in contact with the skin; therefore, prolonged use of the pulse oximeter for neonates or patients with sensitive skin may cause local inflammation or stress due to restricted movement. In addition, owing to COVID-19, there has been a growing demand for the contactless measurement of SpO2. Several studies on measuring SpO2 without contact used skin video images have been conducted. However, in these studies, the SpO2 values were estimated using a linear regression model or a look-up table that required reference values obtained using a contact-type pulse oximeter. In this study, we propose a new technique for the contactless measurement of SpO2 that does not require reference values. Specifically, we used certain approaches that reduced the influence of non-pulsating components and utilized different light wavelengths of video images that penetrated subcutaneously to different depths. We experimentally investigated the accuracy of SpO2 measurements using the proposed methods. The results indicate that the proposed methods were more accurate than the conventional method.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Oxygen SaturationOximetryInfant, NewbornHumansReference ValuesOxygenPulmonary Gas Exchange

Stichwörter der Autor:innen

Facial video imagemultispectral camerapercutaneous oxygen saturationprincipal component analysisvideo plethysmography

Substanzen

Oxygen

Autor:innen und Institutionen

  • Graduate School of EngineeringTohoku University Sendai 9808579 Japan.
  • Cyberscience CenterTohoku University Sendai 9808579 Japan.
  • Center for Promotion of Innovation StrategyTohoku University Sendai 9800845 Japan.

ORCID-Profile: Norihiro Sugita, Makoto Yoshizawa

Förderung

  • Nakatani Foundation for Advancement of Measuring Technologies in Biomedical Engineering

Zitieren

APA

Sasaki, S., Sugita, N., Terai, T., & Yoshizawa, M. (2024). Non-Contact Measurement of Blood Oxygen Saturation Using Facial Video Without Reference Values. IEEE Journal of Translational Engineering in Health and Medicine, 12, 76-83. https://doi.org/10.1109/jtehm.2023.3318643

Vancouver

Sasaki S, Sugita N, Terai T, Yoshizawa M. Non-Contact Measurement of Blood Oxygen Saturation Using Facial Video Without Reference Values. IEEE Journal of Translational Engineering in Health and Medicine. 2024;12:76-83. doi:10.1109/jtehm.2023.3318643

BibTeX
@article{Sasaki2024,
  author = {Sasaki, Soma and Sugita, Norihiro and Terai, Takanori and Yoshizawa, Makoto},
  title = {Non-Contact Measurement of Blood Oxygen Saturation Using Facial Video Without Reference Values},
  journal = {IEEE Journal of Translational Engineering in Health and Medicine},
  year = {2024},
  volume = {12},
  pages = {76--83},
  doi = {10.1109/jtehm.2023.3318643}
}