A Battery-Free Neural-Recording Chip Achieving 5.5 cm Fully-Implanted Depth by Galvanically-Switching Passive Body Channel Communication
IEEE Journal of Solid-State Circuits · 2024-08 · 59 (8) · 2591-2603
DOI: 10.1109/jssc.2024.3366176 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Journal of Solid-State Circuits, 2024-08. Verfasst von Yili Shen und 8 weiteren Autor:innen. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 38 Literaturangaben. Gefördert von: National Key Research and Development Program of China.
Abstract (Originalsprache)
Wireless fully implanted devices are widely adopted for long-term neural-recording applications, where the cable-induced infection risk can be avoided. Battery-free communication based on wireless power transfer (WPT) can eliminate the battery to reduce the size of a wireless implant, realizing minimally invasive surgery. However, conventional battery-free implants suffer from a short communication range, such as inductive coupling, near-infrared (NIR) transmission, and active body-channel communication (BCC), which cannot apply to deep brain zones. Ultrasonic power transfer and communication benefit from a low channel loss, but the low carrier frequency leads to a low data rate, which is not able to transfer full-span neural signals such as spikes and multichannel signals. In this work, a galvanically-switching passive-BCC technique is proposed for neural implants, to extend the effective range of both power transfer and wireless communication. The brain tissue is utilized to form a galvanic loop for power delivery, while the neural-recording data switch the loop current to conduct passive BCC. The proposed technique is implemented in a neural recording chip fabricated in a 55-nm CMOS process. Through-tissue measurement shows that the chip realizes a battery-free communication range of 5.5 cm, with a bit-error rate (BER) of $4.4 \times 10^{-6}$ . In the in-vivo demonstration, a 5.9-mm3 flexible prototype with the proposed chip inside is fully implanted into a Sprague–Dawley rat, where the neural signals are read battery-free through the passive-BCC technique.
Förderung
- National Key Research and Development Program of China
Zitieren
APA
Shen, Y., Yang, C., Zhang, Y., Wang, W., Luo, Y., Yu, C., Xu, K., Pan, G., & Zhao, B. (2024). A Battery-Free Neural-Recording Chip Achieving 5.5 cm Fully-Implanted Depth by Galvanically-Switching Passive Body Channel Communication. IEEE Journal of Solid-State Circuits, 59(8), 2591-2603. https://doi.org/10.1109/jssc.2024.3366176
Vancouver
Shen Y, Yang C, Zhang Y, Wang W, Luo Y, Yu C, et al. A Battery-Free Neural-Recording Chip Achieving 5.5 cm Fully-Implanted Depth by Galvanically-Switching Passive Body Channel Communication. IEEE Journal of Solid-State Circuits. 2024;59(8):2591-2603. doi:10.1109/jssc.2024.3366176
BibTeX
@article{Shen2024,
author = {Shen, Yili and Yang, Changgui and Zhang, Yunshan and Wang, Weixiao and Luo, Yuxuan and Yu, Chaonan and Xu, Kedi and Pan, Gang and Zhao, Bo},
title = {A Battery-Free Neural-Recording Chip Achieving 5.5 cm Fully-Implanted Depth by Galvanically-Switching Passive Body Channel Communication},
journal = {IEEE Journal of Solid-State Circuits},
year = {2024},
volume = {59},
number = {8},
pages = {2591--2603},
doi = {10.1109/jssc.2024.3366176}
}