We report the design, fabrication, and characterization of evanescent mid-infrared germanium-on-silicon waveguide sensors for therapeutic drug monitoring (TDM). TDM requires rapid and accurate quantification of serum drug levels but existing clinical assays rely on laboratory-based instrumentation that limits point-of-care deployment. In this work, tunable diode laser absorption spectroscopy was used to analyze dried samples of the antiseizure medication phenytoin in the spectral region of λ= 5.6–6.0 m. A limit of detection of 2.20 mg/L was achieved for extracted samples, where phenytoin was first added to human serum and subsequently isolated using liquid–liquid extraction. This limit is significantly below the therapeutic window of 10–20 mg/L for phenytoin, enabling detection of sub-therapeutic concentrations. At the same time, the sensor maintains a consistent dose-dependent response up to 40 mg/L, demonstrating its capability to quantify concentrations across the therapeutic window and above the upper therapeutic limit. This validates the use of silicon photonics for biomedical infrared spectroscopy for patients undergoing drug therapy, whether the serum–drug concentration is either too high or too low. These results highlight the potential of mid-IR integrated photonics to form the basis of compact, scalable platforms for point-of-care TDM.
Förderung
- Engineering and Physical Sciences Research Council
Zitieren
APA
Rowe, D. J., Chen, S., Panainte, M., Stirling, C. J., Dong, P., Bakalarz, M., Vizabaskaran, H., Churchill, G., Jin, W., Mourkioti, G., Ebert, M., Reed, G. T., Faust, S. N., Wilkinson, J. S., Nedeljkovic, M., Owens, D. R., & Mashanovich, G. Z. (2026). Mid-Infrared Germanium-on-Silicon Waveguide Sensor for Therapeutic Drug Monitoring of Phenytoins. IEEE Sensors Journal, 26(14), 20840-20849. https://doi.org/10.1109/jsen.2026.3696453
Vancouver
Rowe DJ, Chen S, Panainte M, Stirling CJ, Dong P, Bakalarz M, et al. Mid-Infrared Germanium-on-Silicon Waveguide Sensor for Therapeutic Drug Monitoring of Phenytoins. IEEE Sensors Journal. 2026;26(14):20840-20849. doi:10.1109/jsen.2026.3696453
BibTeX
@article{Rowe2026,
author = {Rowe, David J. and Chen, Siyu and Panainte, Mihai-Adrian and Stirling, Callum J. and Dong, Pin and Bakalarz, Monika and Vizabaskaran, Hanuushah and Churchill, Glenn and Jin, Weilin and Mourkioti, Georgia and Ebert, Martin and Reed, Graham T. and Faust, Saul N. and Wilkinson, James S. and Nedeljkovic, Milos and Owens, Daniel R. and Mashanovich, Goran Z.},
title = {Mid-Infrared Germanium-on-Silicon Waveguide Sensor for Therapeutic Drug Monitoring of Phenytoins},
journal = {IEEE Sensors Journal},
year = {2026},
volume = {26},
number = {14},
pages = {20840--20849},
doi = {10.1109/jsen.2026.3696453}
}
EUVIMED