EUVIMEDEUVIMED entdecken
Crossref

Wearable Organic Electrochemical Transistor System for Multiplexed Chronic-Wound Biosensing

Changyi Zeng, Chenglin Li, Mancheng Li, Chen Chen, Chuan Liu, Songjia Han

IEEE Sensors Journal · 01.02.2026 · 26 (3) · 5098-5105

DOI: 10.1109/jsen.2025.3647151 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Sensors Journal, erschienen am 01.02.2026. Verfasst von Changyi Zeng und 5 weiteren Autor:innen. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 36 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Science and Technology Program of Guangzhou, Natural Science Foundation of Hunan Province.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Chronic nonhealing wounds have emerged as one of the most significant and rapidly expanding clinical complications globally. The accurate, prompt, and early detection of multiple wound-healing biomarkers associated with infection, play a pivotal role in enhancing the treatment and care of such wounds. This study presents an organic electrochemical transistor (OECT)-based wearable bioelectronic system specifically designed for the combined sensing of multiple biomarkers in chronic wounds. The platform integrates an OECT sensor array, a multichannel analog signal processing control circuit, and a mobile application for real-time visualization of detection results. The system enables the simultaneous monitoring of glucose, lactic acid (LA), uric acid (UA), and calcium ion concentrations in wound exudate. This design fills the gap that traditional single-biomarker detection cannot fully reflect the wound healing status. Additionally, by leveraging machine learning algorithms, it effectively mitigates signal crosstalk among different analytes, significantly enhancing the detection accuracy and ensuring the reliability and precision of the monitoring results. This system not only provides multidimensional quantitative evidence for the early detection and prevention of wound infections and deterioration but also offers a proactive approach to managing chronic nonhealing wounds, thereby significantly reducing the associated health risks and improving patient outcomes, providing an innovative paradigm for the clinical translation of multibiomarker applications in the wearable medical field.

Förderung

  • National Natural Science Foundation of China
  • Science and Technology Program of Guangzhou
  • Natural Science Foundation of Hunan Province

Zitieren

APA

Zeng, C., Li, C., Li, M., Chen, C., Liu, C., & Han, S. (2026). Wearable Organic Electrochemical Transistor System for Multiplexed Chronic-Wound Biosensing. IEEE Sensors Journal, 26(3), 5098-5105. https://doi.org/10.1109/jsen.2025.3647151

Vancouver

Zeng C, Li C, Li M, Chen C, Liu C, Han S. Wearable Organic Electrochemical Transistor System for Multiplexed Chronic-Wound Biosensing. IEEE Sensors Journal. 2026;26(3):5098-5105. doi:10.1109/jsen.2025.3647151

BibTeX
@article{Zeng2026,
  author = {Zeng, Changyi and Li, Chenglin and Li, Mancheng and Chen, Chen and Liu, Chuan and Han, Songjia},
  title = {Wearable Organic Electrochemical Transistor System for Multiplexed Chronic-Wound Biosensing},
  journal = {IEEE Sensors Journal},
  year = {2026},
  volume = {26},
  number = {3},
  pages = {5098--5105},
  doi = {10.1109/jsen.2025.3647151}
}