EUVIMEDEUVIMED entdecken
Crossref

Study on the Number and Spacing of Calibration Points for High-Precision NTC Thermistors Across the Ocean Temperature Range

Liquan Sun, Guangyao Wang, Jianping Sun, Weiwei Feng, Hongjun Wang, Ting Li, Lan Du, Hao Huang, Junru Liu

IEEE Sensors Journal · 15.01.2025 · 25 (2) · 2222-2230

DOI: 10.1109/jsen.2024.3510356 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Sensors Journal, erschienen am 15.01.2025. Verfasst von Liquan Sun und 8 weiteren Autor:innen. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 27 Literaturangaben. Gefördert von: National Key Research and Development Program of China, National Natural Science Foundation of China, Fundamental Research Fund.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Marine monitoring often requires the temperature measurement accuracy to be less than 1 mK. A series of regularly arranged temperature points is taken to calibrate the equation expressing resistance-temperature characteristics under ocean conditions. Seventy-two negative temperature coefficient thermistors underwent coincident measurement procedures with nine equally spaced points from -5 °C to 35 °C in a high-precision thermostatic water bath, followed by calibration using Lagrange interpolation and least-squares fitting methods. Ten of these thermistors, with average interpolation errors below 0.1 mK and calibration uncertainty estimated at 0.6 mK (k = 2), were used to study the impact of the number and placement of calibration points on interpolation error, aiming to enhance measurement precision across the ocean temperature range. The results reveal that the standard deviation for the ten thermistors using six unequally spaced points was only 20% higher than when using nine equally spaced points. However, the standard deviation for five equally spaced points nearly doubled, with a maximum difference of 0.1 mK, still achieving sub-millikelvin accuracy. By adjusting the calibration points near both ends of the range, the optimal placement for five calibration points was found to minimize temperature deviation and adhere to the Chebyshev criterion. This ideal distribution, determined through sequential calculations, contributes to improved calibration precision.

Förderung

  • National Key Research and Development Program of China
  • National Natural Science Foundation of China
  • Fundamental Research Fund

Zitieren

APA

Sun, L., Wang, G., Sun, J., Feng, W., Wang, H., Li, T., Du, L., Huang, H., & Liu, J. (2025). Study on the Number and Spacing of Calibration Points for High-Precision NTC Thermistors Across the Ocean Temperature Range. IEEE Sensors Journal, 25(2), 2222-2230. https://doi.org/10.1109/jsen.2024.3510356

Vancouver

Sun L, Wang G, Sun J, Feng W, Wang H, Li T, et al. Study on the Number and Spacing of Calibration Points for High-Precision NTC Thermistors Across the Ocean Temperature Range. IEEE Sensors Journal. 2025;25(2):2222-2230. doi:10.1109/jsen.2024.3510356

BibTeX
@article{Sun2025,
  author = {Sun, Liquan and Wang, Guangyao and Sun, Jianping and Feng, Weiwei and Wang, Hongjun and Li, Ting and Du, Lan and Huang, Hao and Liu, Junru},
  title = {Study on the Number and Spacing of Calibration Points for High-Precision NTC Thermistors Across the Ocean Temperature Range},
  journal = {IEEE Sensors Journal},
  year = {2025},
  volume = {25},
  number = {2},
  pages = {2222--2230},
  doi = {10.1109/jsen.2024.3510356}
}
Kostenlos starten →