EUVIMEDEUVIMED entdecken
Crossref

Electrical Impedance Tomography Based Closed-Loop Tumor Treating Fields in Dynamic Lung Tumors

Maomao Zhang, Yuxi Guo, Xu Xie, Liying Zhu, Yue Lan, Haitang Yang, Yun Pan, Guangdi Chen, Shaomin Zhang, Minmin Wang

IEEE Sensors Journal · 15.12.2024 · 24 (24) · 41989-41996

DOI: 10.1109/jsen.2024.3483558 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Sensors Journal, erschienen am 15.12.2024. Verfasst von Maomao Zhang und 9 weiteren Autor:innen. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 42 Literaturangaben. Gefördert von: China Brain Project, “Pioneer” and “Leading Goose” Research and Development Program of Zhejiang.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Tumor treating fields (TTFields) are a noninvasive anticancer modality that utilizes alternating electric fields (EFs) to disrupt cancer cell division and growth. While generally well-tolerated with minimal side effects, traditional TTFields therapy for lung tumors faces challenges due to the influence of respiratory motion. In this study, we introduce a novel closed-loop TTFields strategy for lung tumors by incorporating electrical impedance tomography (EIT) for real-time respiratory phase monitoring and dynamic parameter adjustments. Furthermore, we conduct theoretical analysis to evaluate the performance of the proposed method using the lung motion model. Compared to conventional TTFields settings, we observed that variations in the electrical conductivity of lung during different respiratory phases led to a decrease in the average EF intensity within lung tumors, transitioning from end-expiratory (1.08 V/cm) to end-inspiratory (0.87 V/cm) phases. Utilizing our proposed closed-loop TTFields approach at the same dose setting (2400 mA, consistent with the traditional TTFields setting), we can achieve a higher and consistent average EF strength at the tumor site (1.30 V/cm) across different respiratory stages. Our proposed closed-loop TTFields method has the potential to improve lung tumor therapy by mitigating the impact of respiratory motion.

Förderung

  • China Brain Project
  • “Pioneer” and “Leading Goose” Research and Development Program of Zhejiang

Zitieren

APA

Zhang, M., Guo, Y., Xie, X., Zhu, L., Lan, Y., Yang, H., Pan, Y., Chen, G., Zhang, S., & Wang, M. (2024). Electrical Impedance Tomography Based Closed-Loop Tumor Treating Fields in Dynamic Lung Tumors. IEEE Sensors Journal, 24(24), 41989-41996. https://doi.org/10.1109/jsen.2024.3483558

Vancouver

Zhang M, Guo Y, Xie X, Zhu L, Lan Y, Yang H, et al. Electrical Impedance Tomography Based Closed-Loop Tumor Treating Fields in Dynamic Lung Tumors. IEEE Sensors Journal. 2024;24(24):41989-41996. doi:10.1109/jsen.2024.3483558

BibTeX
@article{Zhang2024,
  author = {Zhang, Maomao and Guo, Yuxi and Xie, Xu and Zhu, Liying and Lan, Yue and Yang, Haitang and Pan, Yun and Chen, Guangdi and Zhang, Shaomin and Wang, Minmin},
  title = {Electrical Impedance Tomography Based Closed-Loop Tumor Treating Fields in Dynamic Lung Tumors},
  journal = {IEEE Sensors Journal},
  year = {2024},
  volume = {24},
  number = {24},
  pages = {41989--41996},
  doi = {10.1109/jsen.2024.3483558}
}
Kostenlos starten →