Design of Automatic Adjustment Noncontact Sensing Smart Clothing
IEEE Sensors Journal · 15.08.2024 · 24 (16) · 26378-26387
DOI: 10.1109/jsen.2024.3423490 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Sensors Journal, erschienen am 15.08.2024. Verfasst von Heng-Yin Chen und 4 weiteren Autor:innen. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 33 Literaturangaben. Gefördert von: Ministry of Science and Technology, Taiwan, National Science and Technology Council, Taiwan, Higher Education Sprout Project of the National Yang Ming Chiao Tung University and Ministry of Education (MOE), Taiwan.
Abstract (Originalsprache)
Electrocardiogram (ECG), usually measured by Ag/AgCl electrode with conductive gels, is one of the most frequently used tools for the diagnosis of cardiovascular diseases (CVDs) symptoms, but it may encounter the gels drying issue and the risk of skin allergic reaction for long-term ECG monitoring. Recently, the technique of noncontact electrode has developed to access biopotentials without contacting the skin directly to improve the above issue and the convenience and comfort of use, but its signal quality may be affected by the fabric material and the skin-electrode interface condition that still cannot be monitored by current wearable devices. In this study, a novel automatic adjustment noncontact sensing smart clothing was proposed. The designed smart clothing could access biopotentials without contacting the skin directly. Moreover, the function of real-time skin-electrode interface capacitance monitoring was also built to monitor the interface condition simultaneously. The attenuation compensation mechanism was also designed to automatically compensate the attenuation of ECG signal according to the fabric material and the change of interface condition. Finally, the system performance was validated, and the influence of fabric conditions and motion artifact on the change of interface capacitance and ECG signal quality was also investigated. From the experimental results, its ECG signal quality was similar to that of the conventional Ag/AgCl electrodes (correlation >95%) and was stable for long-term monitoring, and the cotton fabric material, thinner fabric thickness, and sweating condition would also improve the ECG signal quality.
Förderung
- Ministry of Science and Technology, Taiwan
- National Science and Technology Council, Taiwan
- Higher Education Sprout Project of the National Yang Ming Chiao Tung University and Ministry of Education (MOE), Taiwan
- Department of Industrial Technology, Ministry of Economic Affairs in Taiwan
- Faculty Group Research Funding Sponsorship through National Taipei University
- “Academic Top-Notch and Features Field Project” Funding Sponsorship of National Taipei University, Taiwan
Zitieren
APA
Chen, H., Lin, B., Yang, S., Chang, W., & Lin, B. (2024). Design of Automatic Adjustment Noncontact Sensing Smart Clothing. IEEE Sensors Journal, 24(16), 26378-26387. https://doi.org/10.1109/jsen.2024.3423490
Vancouver
Chen H, Lin B, Yang S, Chang W, Lin B. Design of Automatic Adjustment Noncontact Sensing Smart Clothing. IEEE Sensors Journal. 2024;24(16):26378-26387. doi:10.1109/jsen.2024.3423490
BibTeX
@article{Chen2024,
author = {Chen, Heng-Yin and Lin, Bor-Shing and Yang, Shih-Rong and Chang, Wei-Ting and Lin, Bor-Shyh},
title = {Design of Automatic Adjustment Noncontact Sensing Smart Clothing},
journal = {IEEE Sensors Journal},
year = {2024},
volume = {24},
number = {16},
pages = {26378--26387},
doi = {10.1109/jsen.2024.3423490}
}