Teleoperated Robot-Assisted Pedicle Tapping: A Hybrid Position–Force Control Approach

Rosaura Morfino, Clemente Lauretti, Francesca Cordella, Loredana Zollo

IEEE Access · 2026 · 14 · 87650-87660

DOI: 10.1109/access.2026.3701980 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Access, 2026. Verfasst von Rosaura Morfino und 3 weiteren Autor:innen. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Italian Ministry of Research, under the Complementary Actions to the NRRP “Fit4MedRob-Fit for Medical Robotics”.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Robotic assistance improves pedicle screw placement accuracy and patient outcomes. However, the tapping phase is still performed manually, introducing outcome variability linked to surgeon expertise. Surgical duration represents another critical factor due to its association with complication rates. Building upon previous work, this study extends a teleoperated semi-autonomous control based on a hybrid position/force approach with the aim of: i) improving screw placement accuracy, and ii) reducing operative time. The proposed control strategy allows the surgeon to switch between different control modes depending on the bone layer: pushing mode or torque mode in the first two bone layers, and a position error based mode in the last layer, where higher accuracy is required. The proposed strategy is evaluated in a controlled feasibility study on a robotic platform by eleven non-expert users performing pedicle tapping on an anthropomorphic spine phantom, with performance of the combined control approach assessed against single-mode operation. Results show that the proposed approach allows an optimal trade-off between procedure effectiveness and efficiency (target error: $TPE = 0.16 \pm 0.11~[\textit{mm}]$ , completion time: $11.77 \pm 2.62 \ [s]$ ), surgeon comfort (RULA score: $2.42 \pm 0.68$ ), and acceptability and usability (NASA-TLX score: $42.1 \pm 13.7$ ). The position error based mode reduces placement error by an order of magnitude compared to other operation modes, while expert feedback confirms moderate workload and preliminary clinical feasibility.

Förderung

  • Italian Ministry of Research, under the Complementary Actions to the NRRP “Fit4MedRob-Fit for Medical Robotics”

Zitieren

APA

Morfino, R., Lauretti, C., Cordella, F., & Zollo, L. (2026). Teleoperated Robot-Assisted Pedicle Tapping: A Hybrid Position–Force Control Approach. IEEE Access, 14, 87650-87660. https://doi.org/10.1109/access.2026.3701980

Vancouver

Morfino R, Lauretti C, Cordella F, Zollo L. Teleoperated Robot-Assisted Pedicle Tapping: A Hybrid Position–Force Control Approach. IEEE Access. 2026;14:87650-87660. doi:10.1109/access.2026.3701980

BibTeX
@article{Morfino2026,
  author = {Morfino, Rosaura and Lauretti, Clemente and Cordella, Francesca and Zollo, Loredana},
  title = {Teleoperated Robot-Assisted Pedicle Tapping: A Hybrid Position–Force Control Approach},
  journal = {IEEE Access},
  year = {2026},
  volume = {14},
  pages = {87650--87660},
  doi = {10.1109/access.2026.3701980}
}
Kostenlos starten →