Improved Speech Recognition with Automated ForwardFocus in Cochlear Kanso 3 Sound Processor Users
Otology & Neurotology Open · 2026-03 · 6 (1) · e087
DOI: 10.1097/ono.0000000000000087 ↗Kurz gefasst
Fachartikel in der Zeitschrift Otology & Neurotology Open, 2026-03. Verfasst von Marian Jones und 3 weiteren Autor:innen. Beteiligte Einrichtung: Cochlear Limited, 1 University Avenue, Macquarie University, NSW, Australia. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Cochlear Limited.
Abstract (Originalsprache)
Objective
The primary objective of this study was to evaluate adult cochlear implant (CI) users’ speech perception in spatially separated noise (S0Nrearhalf) with SCAN 2 FF (automated ForwardFocus) versus SCAN 2 (ForwardFocus off) with the Cochlear Kanso 3 Sound Processor. Secondary objectives included comparing speech perception in the S0N3 configuration, speech perception in quiet between the Kanso 3 Sound Processor and the Kanso 2 Sound Processor, and evaluating user satisfaction and subjective hearing performance with the Kanso 3 Sound Processor versus their own sound processor. Study Design: Single-center, within-subject, interventional study (NCT05619575).
Setting
Purpose-built clinical research unit for auditory device evaluation. Patients: Twenty adults who were postlingually deafened, with at least 6 months of experience with a CI and 3 months of experience with the SCAN program, were recruited.
Intervention
The Kanso 3 Sound Processor with the SCAN 2 FF program. Main Outcome Measure: Paired difference in speech reception threshold (dB SNR) between the SCAN 2 FF and SCAN 2 programs with the Kanso 3 Sound Processor, as measured using the Australian Sentence Test in Noise.
Results
Mean speech recognition thresholds were significantly lower (better) with SCAN 2 FF compared with SCAN 2 in both the S0Nrearhalf and S0N3 configurations by 2.00 dB and 1.43 dB, respectively ( P < 0.001). Speech perception in quiet was comparable between the Kanso 3 and Kanso 2 Sound Processors. Participants reported high satisfaction with the Kanso 3 Sound Processor, and subjective hearing performance was similar to their own sound processor.
Conclusions
This study reinforces the effectiveness of SCAN 2 FF in the Kanso 3 Sound Processor for improving speech perception in noise.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cochlear Limited, 1 University Avenue, Macquarie University, NSW, Australia.
Förderung
- Cochlear Limited
Zitieren
APA
Jones, M., Hong, W., Schou, I. M., & Mashal, M. E. (2026). Improved Speech Recognition with Automated ForwardFocus in Cochlear Kanso 3 Sound Processor Users. Otology & Neurotology Open, 6(1), e087. https://doi.org/10.1097/ono.0000000000000087
Vancouver
Jones M, Hong W, Schou IM, Mashal ME. Improved Speech Recognition with Automated ForwardFocus in Cochlear Kanso 3 Sound Processor Users. Otology & Neurotology Open. 2026;6(1):e087. doi:10.1097/ono.0000000000000087
BibTeX
@article{Jones2026,
author = {Jones, Marian and Hong, Wei and Schou, I Manjula and Mashal, Marjan Emjay},
title = {Improved Speech Recognition with Automated ForwardFocus in Cochlear Kanso 3 Sound Processor Users},
journal = {Otology & Neurotology Open},
year = {2026},
volume = {6},
number = {1},
pages = {e087},
doi = {10.1097/ono.0000000000000087}
}
EUVIMED