The importance of post marketing quality control
European Journal of Anaesthesiology Intensive Care · 2023
DOI: 10.1097/ea9.0000000000000016 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Anaesthesiology Intensive Care, 2023. Verfasst von Philipp K Buehler und 6 weiteren Autor:innen. Beteiligte Einrichtung: From the Department of Anesthesia, University Children Hospital; Zurich, Zurich, Switzerland (PKB, MW, CPB, ARS, MPW, JT), Institute of Intensive Car…. Die Arbeit nennt 6 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
BACKGROUND A previously published study reported on the nonlinear behaviour for airway pressure and tidal volume of the adjustable pressure-limiting (APL) valve of the Aisys CS 2 anaesthesia machine (GE Healthcare, Madison, Wisconsin, USA) during manual bag ventilation. In co-operation with the manufacturer, an adapted APL valve was developed. OBJECTIVE To test and characterise the performance of an adapted APL valve and assess its clinical usability. DESIGN Two-stage study, consisting of an in-vitro testing of an adapted APL valve in a baby and adolescent lung model, followed by a clinical experience survey of anaesthesia personnel after the official implementation of the new APL valve into clinical routine. SETTING Anaesthesia Department at the University Children's Hospital Zurich. MAIN OUTCOME MEASURES Airway pressures and inspiratory tidal volumes during opening and closure of the APL valve at different settings. Likert scale assessment of performance expectance, effort expectance, behavioural intention and safety of the APL valve during clinical use. RESULTS In contrast to the original APL valve of the GE Aisys CS 2 , the adapted APL valve showed a nearly linear increase in airway pressures as well as in the tidal volumes measured. Most importantly, the measured pressures never exceeded the set pressures. Based on the experimental findings, all original APL valves of the GE Aisys CS 2 were replaced by the adapted APL valve. Two months later, an anonymised and standardised questionnaire was handed out to all employees working with the adapted APL valve. Analysis of the questionnaire indicated that the adapted APL valve is easier and more precise to handle in the daily routine than the original one. CONCLUSION The newly adapted APL valve for the GE Aisys CS 2 has considerably improved its linearity at lower pressures and supports our institution's bag mask ventilation practices.
Autor:innen und Institutionen
- From the Department of Anesthesia, University Children Hospital; Zurich, Zurich, Switzerland (PKB, MW, CPB, ARS, MPW, JT), Institute of Intensive Care Medicine, University Hospital Zurich, Zurich, Switzerland (PKB, PDW-…
Zitieren
APA
Buehler, P. K., Weiss, M., Wendel-Garcia, P. D., Both, C. P., Schmidt, A. R., Wehren, M. P., & Thomas, J. (2023). The importance of post marketing quality control. European Journal of Anaesthesiology Intensive Care. https://doi.org/10.1097/ea9.0000000000000016
Vancouver
Buehler PK, Weiss M, Wendel-Garcia PD, Both CP, Schmidt AR, Wehren MP, et al. The importance of post marketing quality control. European Journal of Anaesthesiology Intensive Care. 2023. doi:10.1097/ea9.0000000000000016
BibTeX
@article{Buehler2023,
author = {Buehler, Philipp K and Weiss, Markus and Wendel-Garcia, Pedro D and Both, Christian P and Schmidt, Alexander R and Wehren, Martin P and Thomas, Jörg},
title = {The importance of post marketing quality control},
journal = {European Journal of Anaesthesiology Intensive Care},
year = {2023},
doi = {10.1097/ea9.0000000000000016}
}
EUVIMED