EUVIMEDEUVIMED entdecken
Crossref

Beyond device downloads: residual sleep apnea and hypoxemia missed by flow-based CPAP monitoring

Savit R Malhotra iD, Quynh Theresa Do, Stacey Gunn, Sahil Chopra, Kirstin Knobbe Lloyd, Robert J Thomas iD

SLEEP · 10.11.2026 · 49 (11)

DOI: 10.1093/sleep/zsag189 ↗

Kurz gefasst

Fachartikel in der Zeitschrift SLEEP, online erschienen am 12.07.2026. Verfasst von Savit R Malhotra und 5 weiteren Autor:innen. Beteiligte Einrichtung: Empower Sleep, 530 Technology Drive, Suite 100, Irvine, CA 92618, United States. Hauptthemen laut MEDLINE: Continuous Positive Airway Pressure, Hypoxia, Sleep Apnea, Obstructive, Sleep Apnea Syndromes. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Study Objectives Positive Airway Pressure (PAP) therapy download reports are the current objective data standard for assessing sleep apnea treatment success, providing information on usage, leak, and residual apnea-hypopnea index (AHI). However, follow-up parallel testing can offer added value in measuring sleep and respiratory health in patients receiving PAP therapy. This study compares residual PAP-derived AHI (AHIFLOW) with cardiopulmonary coupling (CPC)-derived AHI3% (AHICPC). We specifically examined patients for whom PAP therapy would be considered effective, but follow-up testing indicated potentially significant residual disease.

Methods

As part of standard clinical care, patients in our telemedicine sleep program undergo longitudinal home sleep testing using CPC through the SleepImage system. We retrospectively analyzed remote physiological monitoring data from ResMed PAP devices and CPC-based sleep testing on nights with concomitant data.

Results

A total of 465 patients met the inclusion criteria, contributing 24 939 nights of matched PAP and CPC data. Mean AHIFLOW was 2.4 events/hour, while mean AHICPC was 12.1 events/hour. Among nights with AHIFLOW < 5 events/hour, 85% of nights had AHICPC ≥ 5 events/hour, 47.7% had AHICPC ≥ 10 events/hour; 28.3% had ≥5 minutes with SpO2 < 90%, and 10.6% had ≥5 minutes with SpO2 < 88%.

Conclusions

PAP download reports are useful for assessing usage and leak, but may underestimate residual respiratory burden. Besides periods of non-use, device-based flow-based metrics alone can miss clinically meaningful residual disease, including hypoxic events during REM sleep. Reliance on PAP-derived AHI as the sole efficacy metric risks overlooking persistent disease.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Continuous Positive Airway PressureHypoxiaSleep Apnea, ObstructiveSleep Apnea SyndromesHumansRetrospective StudiesFemalePolysomnographyMaleMiddle AgedRemote Patient MonitoringAged

Stichwörter der Autor:innen

consumer sleep technologyhome testinghypoxiaobstructive sleep apneapositive airway pressurepulmonary medicinesleep disordered breathing

Autor:innen und Institutionen

  • Empower Sleep, 530 Technology Drive, Suite 100, Irvine, CA 92618, United States.
  • Division of Pulmonary, Critical Care and Sleep Medicine, Department of Medicine, Beth Israel Deaconess Medical Center, Boston, MA 02215, United States.

ORCID-Profile: Savit R Malhotra, Robert J Thomas

Zitieren

APA

Malhotra, S. R., Do, Q. T., Gunn, S., Chopra, S., Knobbe Lloyd, K., & Thomas, R. J. (2026). Beyond device downloads: residual sleep apnea and hypoxemia missed by flow-based CPAP monitoring. SLEEP, 49(11). https://doi.org/10.1093/sleep/zsag189

Vancouver

Malhotra SR, Do QT, Gunn S, Chopra S, Knobbe Lloyd K, Thomas RJ. Beyond device downloads: residual sleep apnea and hypoxemia missed by flow-based CPAP monitoring. SLEEP. 2026;49(11). doi:10.1093/sleep/zsag189

BibTeX
@article{Malhotra2026,
  author = {Malhotra, Savit R and Do, Quynh Theresa and Gunn, Stacey and Chopra, Sahil and Knobbe Lloyd, Kirstin and Thomas, Robert J},
  title = {Beyond device downloads: residual sleep apnea and hypoxemia missed by flow-based CPAP monitoring},
  journal = {SLEEP},
  year = {2026},
  volume = {49},
  number = {11},
  doi = {10.1093/sleep/zsag189}
}
Kostenlos starten →