Automatic alveolar ventilation-targeted NIV mode vs polysomnographic titration of fixed pressure in naive patients with chronic respiratory failure

Mónica Matute-Villacís iD, Marta Puig Dupre, Dunia Gascueña, Anabel Moraleda, Gemma Guerrero Romero, Lidia López-Escuredo, Mireia Dalmases, Àlvar Agustí, Cristina Embid iD

SLEEP · 13.07.2026 · 49 (7)

DOI: 10.1093/sleep/zsag085 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift SLEEP, online erschienen am 26.03.2026. Verfasst von Mónica Matute-Villacís und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pulmonary Medicine, Hospital Universitari Germans Trias i Pujol, Badalona, Spain. Hauptthemen laut MEDLINE: Polysomnography, Respiratory Insufficiency, Noninvasive Ventilation. Die Arbeit nennt 34 Literaturangaben. Gefördert von: Hospital Clínic Barcelona and Societat Catalana de Pneumologia (SOCAP).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Study Objectives Polysomnographic titration of fixed pressures of non-invasive ventilatory support (NIV) is recommended for patients with chronic hypercapnic respiratory failure (CHRF). Here we compared the effects of NIV titrated manually by PSG vs. fully automatic mode (iVAPS AE) in sleep quality, respiratory events manually scored and estimated by built-in software, nocturnal hypoventilation, daytime arterial blood gases and patient comfort.

Methods

We studied 26 adult patients (13 female) with CHRF due to COPD (n = 13), neuromuscular disease (n = 8), obesity hypoventilation syndrome (n = 3) and chest wall disorders (n = 2), all naïve to NIV. They were randomized to standard PSG titration of fixed NIV pressures or PSG using “learn target” iVAPS AE mode in two separate nights, both under PSG monitoring.

Results

Main results showed that:(1) there were no significant differences in the reduction of diurnal PaCO2, nocturnal hypoventilation, sleep quality, upper airway obstruction events with both ventilatory modes; (2) the apnea-hypopnea index from built-in software underestimated that manually scored in both ventilatory modes; (3) ventilatory pressures were lower in fixed mode than iVAPS AE mode; (4) patient comfort was higher in fixed mode than iVAPS AE.

Conclusions

This study shows that iVAPS AE initiated with the “learned target” procedure achieved similar short and long-term ventilatory efficacy to PSG manual bilevel titration. However, its higher pressure levels and lower comfort indicate that the modes are not fully equivalent. Moreover, underestimation of the AHI by the built-in software demands careful interpretation before changing ventilatory settings. Further studies are required to determine the long-term clinical relevance of these differences. Clinical Trial Information This study was registered at ClinicalTrials.gov under the title “Evaluation of Conventional Non-invasive Mechanical Ventilation (NIV) Versus an Automatic Ventilation Mode”. URL: https://clinicaltrials.gov/study/NCT03872167. Registration number: NCT03872167.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PolysomnographyRespiratory InsufficiencyNoninvasive VentilationHumansFemaleMaleMiddle AgedAgedAdultChronic DiseaseObesity Hypoventilation SyndromeSleep Quality

Stichwörter der Autor:innen

iVAPSintelligent volume-assured pressure supportnon invasive ventilationpolysomnographytitration

Autor:innen und Institutionen

  • Department of Pulmonary Medicine, Hospital Universitari Germans Trias i Pujol, Badalona, Spain.
  • Institut de Recerca Germans Trias i Pujol (IGTP), Badalona, Spain.
  • Universitat de Barcelona (UB), Barcelona, Spain.
  • Department of Pulmonary Medicine, Respiratory Institute, Hospital Clinic Barcelona, Barcelona, Spain.
  • Institut d'Investigacions Biomèdiques August Pi i Sunyer (IDIBAPS), Barcelona, Spain.
  • Centro de Investigación Biomédica en Red Enfermedades Respiratorias (CIBERES), Instituto de Salud Carlos III, Madrid, Spain.

ORCID-Profile: Mónica Matute-Villacís, Cristina Embid

Förderung

  • Hospital Clínic Barcelona and Societat Catalana de Pneumologia (SOCAP)

Zitieren

APA

Matute-Villacís, M., Puig Dupre, M., Gascueña, D., Moraleda, A., Guerrero Romero, G., López-Escuredo, L., Dalmases, M., Agustí, À., & Embid, C. (2026). Automatic alveolar ventilation-targeted NIV mode vs polysomnographic titration of fixed pressure in naive patients with chronic respiratory failure. SLEEP, 49(7). https://doi.org/10.1093/sleep/zsag085

Vancouver

Matute-Villacís M, Puig Dupre M, Gascueña D, Moraleda A, Guerrero Romero G, López-Escuredo L, et al. Automatic alveolar ventilation-targeted NIV mode vs polysomnographic titration of fixed pressure in naive patients with chronic respiratory failure. SLEEP. 2026;49(7). doi:10.1093/sleep/zsag085

BibTeX
@article{MatuteVillacs2026,
  author = {Matute-Villacís, Mónica and Puig Dupre, Marta and Gascueña, Dunia and Moraleda, Anabel and Guerrero Romero, Gemma and López-Escuredo, Lidia and Dalmases, Mireia and Agustí, Àlvar and Embid, Cristina},
  title = {Automatic alveolar ventilation-targeted NIV mode vs polysomnographic titration of fixed pressure in naive patients with chronic respiratory failure},
  journal = {SLEEP},
  year = {2026},
  volume = {49},
  number = {7},
  doi = {10.1093/sleep/zsag085}
}
Kostenlos starten →