EUVIMEDEUVIMED entdecken
Multizentrische StudieBeobachtungsstudieOpen Access · CC BY 4.0

Multicenter evaluation of the Waveband system for automated sleep assessment in patients with insomnia symptoms

Silvia Frati Savietto, Antoine Guillot, Mason Harris, Jay Pathmanathan iD, Alexander M Chan iD, M Brandon Westover, Derek Hill, Valérie Bertaina-Anglade, Geoffrey Viardot, Pierrick Arnal, Jacob Donoghue

SLEEP · 13.07.2026 · 49 (7)

DOI: 10.1093/sleep/zsag069 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift SLEEP, online erschienen am 28.03.2026. Verfasst von Silvia Frati Savietto und 10 weiteren Autor:innen. Beteiligte Einrichtung: Beacon Biosignals, Boston, MA, USA. Hauptthemen laut MEDLINE: Electroencephalography, Polysomnography, Sleep Initiation and Maintenance Disorders. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Beacon Biosignals.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Study Objectives This study (Octave-3) aimed to validate the performance of the Waveband dry-electroencephalography sensor device for automatic sleep staging and sleep parameter estimation as compared to gold-standard in-lab polysomnography (PSG).

Methods

Forty-five participants were enrolled, and 38 completed simultaneous PSG and Waveband recordings. PSG data were scored by six human technologists, while Waveband data were scored algorithmically. Agreement between sleep staging results and derived sleep parameters (total sleep time (TST), sleep efficiency (SE), sleep onset latency (SOL), latency to 10 minutes of persistent sleep (LPS), wake after sleep onset (WASO), and time spent in each sleep stage) was measured using the Intraclass Correlation Coefficient (ICC) and overall agreement (OA). Waveband OA was compared to each human rater using the leave-one-out consensus of the remaining five human experts.

Results

Average OA between Waveband vs the leave one out consensuses was 87.3+/-5.4 per cent, equivalent to the average OA for individual human experts of 85.9+/-7.6 per cent (p > .1). Waveband and humans had better OA over the second half of the night, but Waveband had superior OA. ICCs for TST, SE, LPS, and WASO exceeded 0.9, indicating excellent agreement between automated Waveband and human PSG scoring. Lower agreement was found for time spent in N1, N3, and Rapid Eye Movement Sleep (REM), with ICCs ranging from 0.65 to 0.73.

Conclusions

Waveband provides accurate sleep staging and estimation of TST, SE, SOL, LPS, and WASO, with comparable performance to human expert staging of PSG. Its reduced form-factor and good performance should make it a valuable tool for automated assessment of sleep in patients with disturbed sleep. Clinical Trial Information Trial name: Octave-3. URL: https://www.clinicaltrials.gov/study/NCT05438017 Registration: NCT05438017.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectroencephalographyPolysomnographySleep Initiation and Maintenance DisordersAdultFemaleHumansMaleMiddle AgedReproducibility of ResultsSleep DurationSleep Stages

Stichwörter der Autor:innen

Dreem3Sautomatic sleep stagingdry-EEGperformancepolysomnographysleep assessment

Autor:innen und Institutionen

  • Beacon Biosignals, Boston, MA, USA.
  • Panoramic Digital Health, Grenoble, France.
  • Biotrial Neurosciences, Rennes, France.
  • Dreem SAS, Paris, France.

ORCID-Profile: Jay Pathmanathan, Alexander M Chan

Förderung

  • Beacon Biosignals

Zitieren

APA

Savietto, S. F., Guillot, A., Harris, M., Pathmanathan, J., Chan, A. M., Westover, M. B., Hill, D., Bertaina-Anglade, V., Viardot, G., Arnal, P., & Donoghue, J. (2026). Multicenter evaluation of the Waveband system for automated sleep assessment in patients with insomnia symptoms. SLEEP, 49(7). https://doi.org/10.1093/sleep/zsag069

Vancouver

Savietto SF, Guillot A, Harris M, Pathmanathan J, Chan AM, Westover MB, et al. Multicenter evaluation of the Waveband system for automated sleep assessment in patients with insomnia symptoms. SLEEP. 2026;49(7). doi:10.1093/sleep/zsag069

BibTeX
@article{Savietto2026,
  author = {Savietto, Silvia Frati and Guillot, Antoine and Harris, Mason and Pathmanathan, Jay and Chan, Alexander M and Westover, M Brandon and Hill, Derek and Bertaina-Anglade, Valérie and Viardot, Geoffrey and Arnal, Pierrick and Donoghue, Jacob},
  title = {Multicenter evaluation of the Waveband system for automated sleep assessment in patients with insomnia symptoms},
  journal = {SLEEP},
  year = {2026},
  volume = {49},
  number = {7},
  doi = {10.1093/sleep/zsag069}
}