EUVIMEDEUVIMED entdecken
Klinische StudieOpen Access

Wireless wearable sensors can facilitate rapid detection of sleep apnea in hospitalized stroke patients

Jacob Sindorf iD, Alison L Szabo iD, Megan K O'Brien, Aashna Sunderrajan, Kristen L Knutson iD, Phyllis C Zee, Lisa Wolfe, Vineet M Arora, Arun Jayaraman iD

SLEEP · 08.11.2024 · 47 (11)

DOI: 10.1093/sleep/zsae123 ↗

Kurz gefasst

Klinische Studie in der Zeitschrift SLEEP, online erschienen am 30.05.2024. Verfasst von Jacob Sindorf und 8 weiteren Autor:innen. Beteiligte Einrichtung: Max Nader Center for Rehabilitation Technologies and Outcomes Research, Shirley Ryan AbilityLab, Chicago, IL, USA. Hauptthemen laut MEDLINE: Sleep Apnea Syndromes, Stroke, Wearable Electronic Devices. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NICHD NIH HHS, NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Study Objectives To evaluate wearable devices and machine learning for detecting sleep apnea in patients with stroke at an acute inpatient rehabilitation facility (IRF).

Methods

A total of 76 individuals with stroke wore a standard home sleep apnea test (ApneaLink Air), a multimodal, wireless wearable sensor system (ANNE), and a research-grade actigraphy device (ActiWatch) for at least 1 night during their first week after IRF admission as part of a larger clinical trial. Logistic regression algorithms were trained to detect sleep apnea using biometric features obtained from the ANNE sensors and ground truth apnea rating from the ApneaLink Air. Multiple algorithms were evaluated using different sensor combinations and different apnea detection criteria based on the apnea–hypopnea index (AHI ≥ 5, AHI ≥ 15).

Results

Seventy-one (96%) participants wore the ANNE sensors for multiple nights. In contrast, only 48 participants (63%) could be successfully assessed for obstructive sleep apnea by ApneaLink; 28 (37%) refused testing. The best-performing model utilized photoplethysmography (PPG) and finger-temperature features to detect moderate-severe sleep apnea (AHI ≥ 15), with 88% sensitivity and a positive likelihood ratio (LR+) of 44.00. This model was tested on additional nights of ANNE data achieving 71% sensitivity (10.14 LR+) when considering each night independently and 86% accuracy when averaging multi-night predictions.

Conclusions

This research demonstrates the feasibility of accurately detecting moderate-severe sleep apnea early in the stroke recovery process using wearable sensors and machine learning techniques. These findings can inform future efforts to improve early detection for post-stroke sleep disorders, thereby enhancing patient recovery and long-term outcomes. Clinical Trial SIESTA (Sleep of Inpatients: Empower Staff to Act) for Acute Stroke Rehabilitation, https://clinicaltrials.gov/study/NCT04254484?term=SIESTA&checkSpell=false&rank=1, NCT04254484

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Sleep Apnea SyndromesStrokeWearable Electronic DevicesAgedFemaleHumansMaleMiddle AgedActigraphyAlgorithmsHospitalizationMachine LearningPhotoplethysmographyPolysomnographyWireless Technology

Stichwörter der Autor:innen

machine learningobstructive sleep apneasleep apnea screeningstroke

Autor:innen und Institutionen

  • Max Nader Center for Rehabilitation Technologies and Outcomes Research, Shirley Ryan AbilityLab, Chicago, IL, USA.
  • Feinberg School of Medicine, Northwestern University, Chicago, IL, USA.
  • Department of Medicine, University of Chicago Medicine, Chicago, IL, USA.

ORCID-Profile: Jacob Sindorf, Alison L Szabo, Kristen L Knutson, Arun Jayaraman

Förderung

  • NICHD NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Sindorf, J., Szabo, A. L., O'Brien, M. K., Sunderrajan, A., Knutson, K. L., Zee, P. C., Wolfe, L., Arora, V. M., & Jayaraman, A. (2024). Wireless wearable sensors can facilitate rapid detection of sleep apnea in hospitalized stroke patients. SLEEP, 47(11). https://doi.org/10.1093/sleep/zsae123

Vancouver

Sindorf J, Szabo AL, O'Brien MK, Sunderrajan A, Knutson KL, Zee PC, et al. Wireless wearable sensors can facilitate rapid detection of sleep apnea in hospitalized stroke patients. SLEEP. 2024;47(11). doi:10.1093/sleep/zsae123

BibTeX
@article{Sindorf2024,
  author = {Sindorf, Jacob and Szabo, Alison L and O'Brien, Megan K and Sunderrajan, Aashna and Knutson, Kristen L and Zee, Phyllis C and Wolfe, Lisa and Arora, Vineet M and Jayaraman, Arun},
  title = {Wireless wearable sensors can facilitate rapid detection of sleep apnea in hospitalized stroke patients},
  journal = {SLEEP},
  year = {2024},
  volume = {47},
  number = {11},
  doi = {10.1093/sleep/zsae123}
}
Kostenlos starten →