State-of-the-art 32 cm field-of-view digital PET/CT system: preliminary study for protocols optimization and DRLs update

Christian Paolo Luca Fulcheri, Ottavia Sguazzini iD, Veronica D'Ignoti Parenti, Maura Scarlattei, Giorgio Baldari, Veronica Cervati, Tiziano Graziani, Livia Ruffini, Caterina Ghetti iD

Radiation Protection Dosimetry · 02.05.2026 · 202 (7) · 595-608

DOI: 10.1093/rpd/ncag003 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, online erschienen am 02.02.2026. Verfasst von Christian Paolo Luca Fulcheri und 8 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Division, University Hospital of Parma, Via A. Gramsci 14, 43126 Parma, Italy. Hauptthemen laut MEDLINE: Positron Emission Tomography Computed Tomography, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 47 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The aim of this study was to assess the impact of a state-of-the-art 32 cm axial field-of-view GE Omni Legend PET/CT system on administered activity and image quality, and to update Diagnostic Reference Levels (DRLs). System performance was evaluated using NEMA NU 2-2018 measurements and EARL2 accreditation tests. A comparison with the previous 5-ring GE Discovery IQ scanner was performed to analyze the administered activity, scan duration and image quality with and without Artificial Intelligence driven reconstruction. A retrospective analysis of standard-sized adult examinations was conducted to derive updated DRLs values. Omni Legend demonstrated improved sensitivity and Noise Equivalent Count Rate, enabling up to 60% reduction in injected activity while maintaining or improving image quality. Deep learning reconstruction enhanced image quality, particularly at lower administered activities. The DRLs were lower than the national and international reference levels, supporting dose optimization in clinical practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Positron Emission Tomography Computed TomographyImage Processing, Computer-AssistedHumansRadiation DosageRetrospective StudiesDiagnostic Reference LevelsDeep LearningPhantoms, Imaging

Autor:innen und Institutionen

  • Medical Physics Division, University Hospital of Parma, Via A. Gramsci 14, 43126 Parma, Italy.
  • Nuclear Medicine Division, University Hospital of Parma, Via A. Gramsci 14, 43126 Parma, Italy.

ORCID-Profile: Ottavia Sguazzini, Caterina Ghetti

Zitieren

APA

Fulcheri, C. P. L., Sguazzini, O., D'Ignoti Parenti, V., Scarlattei, M., Baldari, G., Cervati, V., Graziani, T., Ruffini, L., & Ghetti, C. (2026). State-of-the-art 32 cm field-of-view digital PET/CT system: preliminary study for protocols optimization and DRLs update. Radiation Protection Dosimetry, 202(7), 595-608. https://doi.org/10.1093/rpd/ncag003

Vancouver

Fulcheri CPL, Sguazzini O, D'Ignoti Parenti V, Scarlattei M, Baldari G, Cervati V, et al. State-of-the-art 32 cm field-of-view digital PET/CT system: preliminary study for protocols optimization and DRLs update. Radiation Protection Dosimetry. 2026;202(7):595-608. doi:10.1093/rpd/ncag003

BibTeX
@article{Fulcheri2026,
  author = {Fulcheri, Christian Paolo Luca and Sguazzini, Ottavia and D'Ignoti Parenti, Veronica and Scarlattei, Maura and Baldari, Giorgio and Cervati, Veronica and Graziani, Tiziano and Ruffini, Livia and Ghetti, Caterina},
  title = {State-of-the-art 32 cm field-of-view digital PET/CT system: preliminary study for protocols optimization and DRLs update},
  journal = {Radiation Protection Dosimetry},
  year = {2026},
  volume = {202},
  number = {7},
  pages = {595--608},
  doi = {10.1093/rpd/ncag003}
}
Kostenlos starten →