EUVIMEDEUVIMED entdecken
Crossref

Prospective dose monitoring using a manual dose management system: experience in brain computed tomography from a tertiary hospital in Nigeria

Idris Garba iD, Engel-Hills Penelope, Florence Davidson, Anas Ismail

Radiation Protection Dosimetry · 2024

DOI: 10.1093/rpd/ncae094 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Idris Garba und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Imaging and Therapeutic Sciences, Faculty of Health and Wellness Sciences, Cape Peninsula University of Technology, Cape Town 8…. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Radiation Dosage, Brain, Tertiary Care Centers. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 36 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A manual radiation dose management system was developed to track the radiation dose and scan parameters of patients for brain computed tomography (CT). Radiation dose in volume computed tomography dose index (CTDIvol) and dose length product (DLP) were monitored to identify procedures that may require optimisation using notification values. The data were analysed and compared with national and international diagnostic reference levels (DRLs). A total of 596 brain CTs were monitored and grouped as <1: 36, 1–<5: 38, 5–<10: 25, 10–<15: 31 and adult: 466. The CTDIvol notification value identified the following number of examinations having high CTDIvol in <1 y: 1, 1–<5: 1, 5–<10: 0, 10–<15: 0 and adult (>15): 11. Furthermore, the DLP notification values identified the following examinations with high DLP in <1 y: 1, 1–<5:1, 5–<10:1, 10–<15: 1 and adults (>15): 18. The established local paediatric DLP DRLs were 2–3 times higher than the international paediatric DLP DRLs. This calls for a total protocol review and optimisation considering the local CT practices for paediatric imaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedRadiation DosageBrainTertiary Care CentersRadiation MonitoringHumansNigeriaChildAdultChild, PreschoolFemaleProspective StudiesMaleAdolescentInfantYoung AdultMiddle AgedAgedInfant, NewbornRadiation Protection

Autor:innen und Institutionen

  • Department of Medical Imaging and Therapeutic Sciences, Faculty of Health and Wellness Sciences, Cape Peninsula University of Technology, Cape Town 8000, South Africa.
  • Department of Radiology, Faculty of Clinical Sciences, College of Health Sciences, Bayero University Kano, Kano 700001, Nigeria.

ORCID-Profile: Idris Garba

Zitieren

APA

Garba, I., Penelope, E. H., Davidson, F., & Ismail, A. (2024). Prospective dose monitoring using a manual dose management system: experience in brain computed tomography from a tertiary hospital in Nigeria. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncae094

Vancouver

Garba I, Penelope EH, Davidson F, Ismail A. Prospective dose monitoring using a manual dose management system: experience in brain computed tomography from a tertiary hospital in Nigeria. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncae094

BibTeX
@article{Garba2024,
  author = {Garba, Idris and Penelope, Engel-Hills and Davidson, Florence and Ismail, Anas},
  title = {Prospective dose monitoring using a manual dose management system: experience in brain computed tomography from a tertiary hospital in Nigeria},
  journal = {Radiation Protection Dosimetry},
  year = {2024},
  doi = {10.1093/rpd/ncae094}
}
Kostenlos starten →