High radon levels and adverse environmental conditions in using RAD7
Radiation Protection Dosimetry · 02.11.2023 · 199 (18) · 2199-2202
DOI: 10.1093/rpd/ncad205 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Protection Dosimetry, online erschienen am 03.11.2023. Verfasst von Arturo Lopez und 2 weiteren Autor:innen. Beteiligte Einrichtung: Instituto Nacional de Investigaciones Nucleares, Carretera México Toluca S/N, La Marquesa C.P. 52150, Ocoyoacac, Estado de Mexico, Mexico. Hauptthemen laut MEDLINE: Radon, Uranium, Radiation Monitoring, Air Pollutants, Radioactive. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 5 Literaturangaben. Gefördert von: Institut Pluridisciplinaire Hubert Curien UMR 7178 CNRS/Strasbourg University.
Abstract (Originalsprache)
Good response was observed in simultaneously using six RAD7 detectors at high radon concentrations, temperatures and relative humidity conditions. RAD7 detectors were tested in laboratory using radon chambers from 13 up to 59.8 kBq m−3, statistical analysis allows to distinct between counting errors and radon variation. RAD7 detectors were exposure to extreme environmental conditions at uranium flat mine. High radon concentration in soil from a confined uranium mineral was 274.3 kBq m−3 at 44°C temperature and 20% relative humidity. Inside uranium mine radon increased from 1.0 up to 88.5 kBq m−3.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Instituto Nacional de Investigaciones Nucleares, Carretera México Toluca S/N, La Marquesa C.P. 52150, Ocoyoacac, Estado de Mexico, Mexico.
- Servicio Geológico Mexicano, Blvd. Felipe Ángeles km. 93.5-4, Col. Venta Prieta C.P. 42080, Pachuca de Soto, Hidalgo, México.
Förderung
- Institut Pluridisciplinaire Hubert Curien UMR 7178 CNRS/Strasbourg University
Zitieren
APA
Lopez, A., Balcazar, M., & Sánchez, D. (2023). High radon levels and adverse environmental conditions in using RAD7. Radiation Protection Dosimetry, 199(18), 2199-2202. https://doi.org/10.1093/rpd/ncad205
Vancouver
Lopez A, Balcazar M, Sánchez D. High radon levels and adverse environmental conditions in using RAD7. Radiation Protection Dosimetry. 2023;199(18):2199-2202. doi:10.1093/rpd/ncad205
BibTeX
@article{Lopez2023,
author = {Lopez, Arturo and Balcazar, Miguel and Sánchez, David},
title = {High radon levels and adverse environmental conditions in using RAD7},
journal = {Radiation Protection Dosimetry},
year = {2023},
volume = {199},
number = {18},
pages = {2199--2202},
doi = {10.1093/rpd/ncad205}
}
EUVIMED