EUVIMEDEUVIMED entdecken
Open Access

Automation of Functional Mobility Assessments at Home Using a Multimodal Sensor System Integrating Inertial Measurement Units and Computer Vision (IMU-Vision)

Automatisierung von funktionellen Mobilitätsassessments zu Hause mithilfe eines multimodalen Sensorsystems mit Inertialmesssensoren und Computer Vision (IMU-Vision)

Johanna Spangler, Marc Mitjans, Ashley Collimore, Aysha Gomes-Pires, David M Levine, Roberto Tron, Louis N Awad iD

Physical Therapy · 01.02.2024 · 104 (2)

DOI: 10.1093/ptj/pzad184 ↗

Auf einen Blick

Bei 50 frail und prefrail adults zeigte das IMU-Vision-System im Home-Setting gute Durchführbarkeit und hohe Genauigkeit bei der Erkennung und Bewertung funktioneller Aktivitäten.

Fragestellung

Lässt sich ein multimodales Bewegungserfassungssystem zur autonomen, häuslichen Erkennung und Bewertung funktioneller Mobilitätsaufgaben einsetzen?

Methode

Studie bei 50 frail und prefrail adults aus dem Brigham and Women’s Hospital at Home program; Vergleich manueller mit automatischer Erkennung und Bewertung von Sit-to-Stand, Gleichgewicht und Gehen zu Hause.

Ergebnisse

Von 271 Testaktivitäten waren 217 (80 %) analysierbar; 206 (95 %) der analysierbaren Aktivitäten wurden korrekt erkannt. Gehen: 100 %, Gleichgewicht: 97 %, Sit-to-Stand: 82 %; bei 152 Tests stimmte die automatische mit der manuellen Bewertung gut überein (Kw = 0,76 [0,69, 0,83]).

Einschränkungen

Für die automatische Bewertung waren nur 152 Tests geeignet; eine weitere Algorithmustrainierung mit ökologisch validen Daten ist nötig.

Schlussfolgerung der Autor:innen

Die autonome Erkennung und Bewertung häuslicher funktioneller Aktivitäten wird durch ein multimodales System mit IMU- und Visionsdaten ermöglicht; weitere Training und ein selbstständig nutzbares System sind vor vollständiger Autonomie erforderlich.

Kernaussagen

  • 217/271 Tests waren analysierbar
  • 206/217 analysierbare Aktivitäten wurden korrekt erkannt
  • Automatische und manuelle Scores: Kw = 0,76 [0,69, 0,83]

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Physical Therapy, online erschienen am 30.12.2023. Verfasst von Johanna Spangler und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physical Therapy, Sargent College of Health and Rehabilitation Sciences, Boston University, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Wearable Electronic Devices. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIA NIH HHS, NIH HHS, National Institutes of Health.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Functional movement assessments are routinely used to evaluate and track changes in mobility. The objective of this study was to evaluate a multimodal movement monitoring system developed for autonomous, home-based, functional movement assessment.

Methods

Fifty frail and prefrail adults were recruited from the Brigham and Women’s Hospital at Home program to evaluate the feasibility and accuracy of applying the multimodal movement monitoring system to autonomously recognize and score functional activities collected in the home. Study subjects completed sit-to-stand, standing balance (Romberg, semitandem, and tandem), and walking test activities in likeness to the Short Physical Performance Battery. Test activities were identified and scored manually and by the multimodal movement monitoring system’s activity recognition and scoring algorithms, which were previously trained on lab-based biomechanical data to integrate wearable inertial measurement unit (IMU) and external red–blue–green-depth vision data. Feasibility was quantified as the proportion of completed tests that were analyzable. Accuracy was quantified as the degree of agreement between the actual and system-identified activities. In an exploratory analysis of a subset of functional activity data, the accuracy of a preliminary activity-scoring algorithm was also evaluated.

Results

Activity recognition by the IMU-vision system had good feasibility and high accuracy. Of 271 test activities collected in the home, 217 (80%) were analyzable by the activity-recognition algorithm, which overall correctly identified 206 (95%) of the analyzable activities: 100% of walking, 97% of balance, and 82% of sit-to-stand activities (χ2(2) = 19.9). In the subset of 152 tests suitable for activity scoring, automatic and manual scores showed substantial agreement (Kw = 0.76 [0.69, 0.83]).

Conclusions

Autonomous recognition and scoring of home-based functional activities is enabled by a multimodal movement monitoring system that integrates inertial measurement unit and vision data. Further algorithm training with ecologically valid data and a kitted system that is independently usable by patients are needed before fully autonomous, functional movement assessment is realizable. Impact Functional movement assessments that can be administered in the home without a clinician present have the potential to democratize these evaluations and improve care access.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Wearable Electronic DevicesAdultHumansFemaleMovementWalkingAutomationComputers

Stichwörter der Autor:innen

AutomatedComputer VisionFunctional Movement AssessmentHomeIMUsWearable

Autor:innen und Institutionen

  • Department of Physical Therapy, Sargent College of Health and Rehabilitation Sciences, Boston University, Boston, Massachusetts, USA.
  • Department of Mechanical Engineering, College of Engineering, Boston University, Boston, Massachusetts, USA.
  • Division of General Internal Medicine and Primary Care, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.

ORCID-Profile: Louis N Awad

Förderung

  • NIA NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Spangler, J., Mitjans, M., Collimore, A., Gomes-Pires, A., Levine, D. M., Tron, R., & Awad, L. N. (2024). Automation of Functional Mobility Assessments at Home Using a Multimodal Sensor System Integrating Inertial Measurement Units and Computer Vision (IMU-Vision). Physical Therapy, 104(2). https://doi.org/10.1093/ptj/pzad184

Vancouver

Spangler J, Mitjans M, Collimore A, Gomes-Pires A, Levine DM, Tron R, et al. Automation of Functional Mobility Assessments at Home Using a Multimodal Sensor System Integrating Inertial Measurement Units and Computer Vision (IMU-Vision). Physical Therapy. 2024;104(2). doi:10.1093/ptj/pzad184

BibTeX
@article{Spangler2024,
  author = {Spangler, Johanna and Mitjans, Marc and Collimore, Ashley and Gomes-Pires, Aysha and Levine, David M and Tron, Roberto and Awad, Louis N},
  title = {Automation of Functional Mobility Assessments at Home Using a Multimodal Sensor System Integrating Inertial Measurement Units and Computer Vision (IMU-Vision)},
  journal = {Physical Therapy},
  year = {2024},
  volume = {104},
  number = {2},
  doi = {10.1093/ptj/pzad184}
}
Kostenlos starten →