Dermabond versus suture fixation for spinal cord stimulator trial lead anchoring: a randomized within-patient controlled trial
Pain Medicine · 01.11.2026 · 27 (11) · 1221-1230
DOI: 10.1093/pm/pnag083 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Pain Medicine, online erschienen am 03.07.2026. Verfasst von Robert W Hurley und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Wake Forest University School of Medicine, Medical Center Boulevard, Winston-Salem, NC, United States. Hauptthemen laut MEDLINE: Spinal Cord Stimulation, Cyanoacrylates, Tissue Adhesives, Suture Techniques. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Objective
To determine whether Dermabond (2-octyl cyanoacrylate) provides superior lead stability to suture for percutaneous spinal cord stimulator (SCS) trial leads, and whether migration reflects patient-level factors rather than anchoring method.
Design
Prospective, randomized, within-patient controlled trial (ClinicalTrials.gov NCT05914311).
Setting
Interventional Pain and Spine Medicine Center at an academic medical center. Subjects Forty-eight adults (96 leads) undergoing bilateral thoracic percutaneous SCS trial implantation.
Methods
Within each patient, one lead was randomized to Dermabond and the contralateral lead to suture. The primary outcome was absolute radiographic lead migration (mm) between placement and end-of-trial, both imaged upright. All analyses accounted for within-patient correlation; a post hoc non-inferiority framing used the a priori 9 mm clinically meaningful threshold.
Results
Mean absolute migration was 9.15 ± 14.49 mm (Dermabond) and 10.60 ± 17.42 mm (suture); the paired difference was −1.46 mm (95% CI, −6.73 to 3.81; P = .58), with the upper bound well below the 9 mm threshold. In adjusted analyses, male sex (+8.17 mm; P = .036) and prior spine surgery (+7.08 mm; P = .048) predicted greater migration; body mass index (BMI), age, and fixation method did not. Clinically significant (≥9 mm) migration occurred in 27.1% of Dermabond and 22.9% of suture leads (McNemar P = .77).
Conclusions
Dermabond was not superior to suture but showed clinically comparable stability, supporting it as a noninvasive alternative. The stronger signal was at the patient level: male sex and prior spine surgery, but not BMI, predicted migration. Future efforts to reduce migration should target patient-level drivers rather than anchor choice. Clinical trial registration number ClinicalTrials.gov NCT05914311 (registered a priori).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Anesthesiology, Wake Forest University School of Medicine, Medical Center Boulevard, Winston-Salem, NC, United States.
- Department of Translational Neuroscience, Wake Forest University School of Medicine, Medical Center Boulevard, Winston-Salem, NC, United States.
- Department of Public Health, Wake Forest University School of Medicine, Medical Center Boulevard, Winston-Salem, NC, United States.
- Pain Outcomes Lab, Wake Forest University Health Sciences, Medical Center Boulevard, Winston-Salem, NC, United States.
ORCID-Profile: Robert W Hurley
Zitieren
APA
Hurley, R. W., Patel, J. S., Bintrim, D. J., Oliver, J. B., Miranda, A., Rahmah, T., & Hamsher, C. P. (2026). Dermabond versus suture fixation for spinal cord stimulator trial lead anchoring: a randomized within-patient controlled trial. Pain Medicine, 27(11), 1221-1230. https://doi.org/10.1093/pm/pnag083
Vancouver
Hurley RW, Patel JS, Bintrim DJ, Oliver JB, Miranda A, Rahmah T, et al. Dermabond versus suture fixation for spinal cord stimulator trial lead anchoring: a randomized within-patient controlled trial. Pain Medicine. 2026;27(11):1221-1230. doi:10.1093/pm/pnag083
BibTeX
@article{Hurley2026,
author = {Hurley, Robert W and Patel, Janus S and Bintrim, Daniel J and Oliver, Jennifer B and Miranda, Amelia and Rahmah, Tiye and Hamsher, Carlyle P},
title = {Dermabond versus suture fixation for spinal cord stimulator trial lead anchoring: a randomized within-patient controlled trial},
journal = {Pain Medicine},
year = {2026},
volume = {27},
number = {11},
pages = {1221--1230},
doi = {10.1093/pm/pnag083}
}