A novel low-profile self-expanding epidural lead array system that is fully collapsible, deployable, and retrievable
Pain Medicine · 2025
DOI: 10.1093/pm/pnaf124 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pain Medicine, 2025. Verfasst von Hassan Beheshti Seresht und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Industrial Engineering, University of Pittsburgh, Pittsburgh, PA 15260, United States. Hauptthemen laut MEDLINE: Electrodes, Implanted, Spinal Cord Stimulation, Epidural Space, Pain Management. Die Arbeit nennt 34 Literaturangaben. Gefördert von: Center for Medical Innovation, University of Pittsburgh.
Abstract (Originalsprache)
Objective
Introduced in 1970s, Spinal Cord Stimulator (SCS) devices have played a crucial role in managing a wide range of complex and refractory chronic pain, particularly back/leg pain as well as neuropathic pain. Currently, 2 primary types of leads, cylindrical and paddle leads, are prevalent in pain management. While both effectively alleviate pain, cylindrical leads, due to their small size, are susceptible to movement and migration as well as a smaller surface area for coverage, leading to device displacement and failure to provide pain relief. On the other hand, paddle leads offer a larger surface area and secure placement but require a relatively large incision for device insertion. To address the limitations of existing SCS devices, a novel SCS device has been developed with a low-profile, deployable, and retrievable design based on the human epidural anatomy. Methods and Results A prototype SCS has been successfully designed, fabricated, and tested in vitro. This innovative design features a laser-trimmed nitinol mesh structure as the self-expanding deployable frame, an ultrathin ePTFE membrane isolating the conductive metallic frame, and platinum-iridium materials which can be seamlessly integrate with an external battery pack for the delivery of efficient electrical potential. The anatomical nitinol mesh frame allows the entire device to collapse into a size smaller than that of a 14-gauge needle (1.5-1.6 mm in diameter) and its corresponding delivery sheath. Additionally, mechanical and electrochemical tests were carried out to assess the performance of the developed device. The mechanical tests demonstrated the backbone’s ability to expand within the epidural space. Similarly, electrochemical tests on the electrodes underscored that the selected materials were indeed appropriate.
Conclusion
This novel SCS design effectively prevents device dislocation and migration showing great wall apposition while providing a larger surface area for pain management. These results support its potential as a next-generation platform for effective chronic pain management.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Industrial Engineering, University of Pittsburgh, Pittsburgh, PA 15260, United States.
- Department of Anesthesiology and Perioperative Medicine, Division of Chronic Pain, University of Pittsburgh Medical Center, Pittsburgh, PA 15260, United States.
- Department of Mechanical Engineering and Material Science, University of Pittsburgh, Pittsburgh, PA 15260, United States.
- Department of Bioengineering, University of Pittsburgh, Pittsburgh, PA 15260, United States.
- McGowan Institute for Regenerative Medicine, University of Pittsburgh, Pittsburgh, PA 15260, United States.
ORCID-Profile: Hassan Beheshti Seresht, Trent D Emerick
Förderung
- Center for Medical Innovation
- University of Pittsburgh
Zitieren
APA
Beheshti Seresht, H., Emerick, T. D., Chauhan, G., Ibrahim, M. S., Jung, Y., Lee, J. K., & Chun, Y. (2025). A novel low-profile self-expanding epidural lead array system that is fully collapsible, deployable, and retrievable. Pain Medicine. https://doi.org/10.1093/pm/pnaf124
Vancouver
Beheshti Seresht H, Emerick TD, Chauhan G, Ibrahim MS, Jung Y, Lee JK, et al. A novel low-profile self-expanding epidural lead array system that is fully collapsible, deployable, and retrievable. Pain Medicine. 2025. doi:10.1093/pm/pnaf124
BibTeX
@article{BeheshtiSeresht2025,
author = {Beheshti Seresht, Hassan and Emerick, Trent D and Chauhan, Gaurav and Ibrahim, Mohamed S and Jung, Youngsoo and Lee, Jung-Kun and Chun, Youngjae},
title = {A novel low-profile self-expanding epidural lead array system that is fully collapsible, deployable, and retrievable},
journal = {Pain Medicine},
year = {2025},
doi = {10.1093/pm/pnaf124}
}
EUVIMED