Accuracy and Usability of a New Rapid Diagnostic Test for the Diagnosis of Dengue in Vientiane, Laos
Open Forum Infectious Diseases · 01.07.2026 · 13 (7)
DOI: 10.1093/ofid/ofag422 ↗Kurz gefasst
Fachartikel in der Zeitschrift Open Forum Infectious Diseases, online erschienen am 14.07.2026. Verfasst von Vilayouth Phimolsannousith und 10 weiteren Autor:innen. Beteiligte Einrichtung: Lao-Oxford-Mahosot Hospital-Wellcome Trust Research Unit, Microbiology Laboratory, Mahosot Hospital, Vientiane, Lao People's Democratic Republic. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome Trust, Institute of Research for Development, Aix-Marseille University.
Abstract (Originalsprache)
Background
An estimated 390 million dengue virus (DENV) infections occur per year, 70% in the Western Pacific and Southeast Asia. A prototype rapid diagnostic test (RDT) (NS1, immunoglobulin M [IgM]/immunoglobulin G [IgG]; DengueDx, Mologic) was evaluated for accuracy and ease of use for dengue diagnosis at Mahosot Hospital, Vientiane, Lao People’s Democratic Republic.
Methods
Five hundred twenty-six sera collected in 2017 and 2018 from patients hospitalized with suspected dengue were included. Prototype RDT results were compared to NS1 and IgM enzyme-linked immunosorbent assays and/or real-time reverse-transcription polymerase chain reaction as reference assays and to Bioline dengue RDT (NS1, IgM/IgG) as comparator. The accuracy of the image-based result interpretation (RDT scan or photograph) at day 0 and after a delay of 7 days was evaluated. Five technicians were observed when performing RDTs and interviewed on their perception of the ease of use.
Results
The overall sensitivity and the sensitivity for NS1 detection alone were significantly lower for the prototype RDT (65.2% [95% confidence interval {CI}, 59.3%–70.8%] and 44.6% [95% CI, 38.0%–51.4%], respectively) than for the Bioline RDT (76.7% [95% CI, 69.1%–83.2%], z-test P = .014; and 66.7% [95% CI, 57.5%–75.0%], z-test P < .001, respectively). The positive predictive value of the prototype RDT for detection of NS1 was >90%. Only half of the positive results were found positive when reading results from images. Correlation of >98% for positive results was obtained between reading at day 0 and reading at day 7. During ease of use assessment, when performing the Bioline RDTs the participants made fewer errors and were more satisfied than when performing the prototype RDTs.
Conclusions
Improvement in test sensitivity and in device design would adapt the prototype RDT for Laos and similar contexts. Optimization of the prototype RDT is warranted and further evaluation in different healthcare settings would provide a more accurate estimation of the value of this RDT for dengue surveillance in remote areas.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Lao-Oxford-Mahosot Hospital-Wellcome Trust Research Unit, Microbiology Laboratory, Mahosot Hospital, Vientiane, Lao People's Democratic Republic.
- Centre for Tropical Medicine and Global Health, Nuffield Department of Clinical Medicine, University of Oxford, Oxford, United Kingdom.
- Institute of Research and Education Development, University of Health Sciences, Vientiane, Lao People's Democratic Republic.
- Saw Swee Hock School of Public Health, National University of Singapore, Singapore, Singapore.
- Mahidol Oxford Tropical Medicine Research Unit, Faculty of Tropical Medicine, Mahidol University, Bangkok, Thailand.
- Faculty of Infectious and Tropical Diseases, London School of Hygiene and Tropical Medicine, London, United Kingdom.
ORCID-Profile: Mayfong Mayxay, Heidi Hopkins, Audrey Dubot-Pérès
Förderung
- Wellcome Trust
- Institute of Research for Development
- Aix-Marseille University
- Global Access Dx
Zitieren
APA
Phimolsannousith, V., Sibounheuang, B., Chanthongthip, A., Vongsouvath, M., Mayxay, M., Mukaka, M., Luangraj, M., Newton, P. N., Ashley, E. A., Hopkins, H., & Dubot-Pérès, A. (2026). Accuracy and Usability of a New Rapid Diagnostic Test for the Diagnosis of Dengue in Vientiane, Laos. Open Forum Infectious Diseases, 13(7). https://doi.org/10.1093/ofid/ofag422
Vancouver
Phimolsannousith V, Sibounheuang B, Chanthongthip A, Vongsouvath M, Mayxay M, Mukaka M, et al. Accuracy and Usability of a New Rapid Diagnostic Test for the Diagnosis of Dengue in Vientiane, Laos. Open Forum Infectious Diseases. 2026;13(7). doi:10.1093/ofid/ofag422
BibTeX
@article{Phimolsannousith2026,
author = {Phimolsannousith, Vilayouth and Sibounheuang, Bountoy and Chanthongthip, Anisone and Vongsouvath, Manivanh and Mayxay, Mayfong and Mukaka, Mavuto and Luangraj, Manophab and Newton, Paul N and Ashley, Elizabeth A and Hopkins, Heidi and Dubot-Pérès, Audrey},
title = {Accuracy and Usability of a New Rapid Diagnostic Test for the Diagnosis of Dengue in Vientiane, Laos},
journal = {Open Forum Infectious Diseases},
year = {2026},
volume = {13},
number = {7},
doi = {10.1093/ofid/ofag422}
}
EUVIMED