Studienprotokoll

Monitoring Respiratory Rate Continuously Without Attaching a Sensor During a Challenging Ramped Protocol

Ulf R Borg iD, André Antunes, Philip Smit, Paul S Addison, Dean Montgomery

Military Medicine · 2024

DOI: 10.1093/milmed/usae200 ↗

Kurz gefasst

Studienprotokoll in der Zeitschrift Military Medicine, 2024. Verfasst von Ulf R Borg und 4 weiteren Autor:innen. Beteiligte Einrichtung: Medical Affairs, Medtronic Patient Monitoring, Boulder, CO 80301, USA. Hauptthemen laut MEDLINE: Respiratory Rate. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Gefördert von: Medtronic.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Respiratory rate (RR) is a crucial vital sign in patient monitoring and is often the best marker of the deterioration of a sick patient. It can be used to help diagnose numerous medical conditions and has been demonstrated to be an independent predictor of patient outcomes in various critical care settings and is incorporated in many clinical early warning scores. Here, we report on the performance of depth-camera-based system for the noncontact monitoring of RR during a ramped RR protocol. The ramped breathing protocol was developed specifically to test the relatively rapid changes in rates, which include clinically important low and high ranges of RRs. Materials and Methods We performed a series of experimental runs with healthy volunteers who were instructed to breathe over a wide range of RRs, where the rates were ramped up from 4 breaths/min to 50 breaths/min then back down to 4 breaths/min in a series of ramped steps. Depth information was acquired from the scene and used to determine a respiratory rate (RRdepth), and this was compared to capnograph or spirometer respiratory rate reference (RRref). A total of 9,482 contemporaneous data pairs (RRdepth, RRref) were collected during the study for comparison.

Results

A Pearson correlation coefficient of 0.995 was achieved and a line of best fit given by RRdepth = 0.99 × RRref + 0.36 breaths/min. The overall root mean squared difference (RMSD) across the runs was 1.29 breaths/min with a corresponding bias of 0.16 breaths/min, respectively. The associated Bland–Altman analysis found limits of agreement of −2.45 and 2.75 breaths/min. When the data were subdivided according to low, medium, and high RRs, corresponding to ≤10, >10 to 20, and >20 breaths/min, the RMSD accuracies were found to be 0.94, 1.34, and 1.55 breaths/min, respectively.

Conclusions

The technology performed well, exhibiting an RMSD accuracy well within our target of 3 breaths/min, both across the whole range and across each individual subrange. In summary, our results indicate the potential viability of continuous noncontact monitoring for the determination of RR over a clinically relevant range.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Respiratory RateHumansMaleMonitoring, PhysiologicAdultFemaleHealthy Volunteers

Autor:innen und Institutionen

  • Medical Affairs, Medtronic Patient Monitoring, Boulder, CO 80301, USA.
  • Medtronic Patient Monitoring, Technopole Centre, Edinburgh, UK.

ORCID-Profile: Ulf R Borg

Förderung

  • Medtronic

Zitieren

APA

Borg, U. R., Antunes, A., Smit, P., Addison, P. S., & Montgomery, D. (2024). Monitoring Respiratory Rate Continuously Without Attaching a Sensor During a Challenging Ramped Protocol. Military Medicine. https://doi.org/10.1093/milmed/usae200

Vancouver

Borg UR, Antunes A, Smit P, Addison PS, Montgomery D. Monitoring Respiratory Rate Continuously Without Attaching a Sensor During a Challenging Ramped Protocol. Military Medicine. 2024. doi:10.1093/milmed/usae200

BibTeX
@article{Borg2024,
  author = {Borg, Ulf R and Antunes, André and Smit, Philip and Addison, Paul S and Montgomery, Dean},
  title = {Monitoring Respiratory Rate Continuously Without Attaching a Sensor During a Challenging Ramped Protocol},
  journal = {Military Medicine},
  year = {2024},
  doi = {10.1093/milmed/usae200}
}
Kostenlos starten →