Sigma metric and Quality Goal Index assessment of hemoglobin A1c analytical performance in an ambulatory accredited laboratory

Jagdish Chandarana iD, Kinjal Butani iD

Laboratory Medicine · 04.08.2026 · 57 (5)

DOI: 10.1093/labmed/lmag052 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Laboratory Medicine, online erschienen am 20.08.2026. Verfasst von Jagdish Chandarana und 1 weiteren Autor:innen. Beteiligte Einrichtung: Vedant Diagnostic Center, Rajkot, Gujarat, India. Hauptthemen laut MEDLINE: Glycated Hemoglobin, Laboratories, Clinical. Die Arbeit nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Hemoglobin A1c has a stringent total allowable error (TEa), making high Six Sigma values difficult to attain. In addition, data from ambulatory accredited laboratories are sparse.

Methods

Internal quality control (QC) data for 2 levels were collected over 8 months on a single-lot Bio-Rad D-10 high-performance liquid chromatography analyzer. Long-term imprecision was the coefficient of variation (s/x¯) × 100 (CV) of all internal QC results combined, capturing between-month variation. Bias was derived from external quality assessment (EQAS) and assigned QC targets, applied as absolute values. The Sigma metric was (TEa ‒ absolute bias) / CV at TEa levels of 6%, 7%, and 8%; the Quality Goal Index distinguished imprecision-driven from bias-driven performance.

Results

Long-term CVs were 3.04% (level 1) and 1.34% (level 2); EQASs were satisfactory throughout. At a TEa of 6% with EQAS-derived bias, Sigma metrics were 1.19 and 2.70 and, with a level-specific QC target bias, 1.46 and 4.22. At the Clinical Laboratory Improvement Amendments 2025 criterion of 8%, the Sigma metrics were 2.11 and 5.71. The level 1 Quality Goal Index (0.52) identified imprecision as predominant.

Discussion

Despite satisfactory external proficiency, Sigma metrics fell below 3 at both levels against a 6% TEa and below 2 at the clinically pivotal lower concentration, driven by imprecision rather than bias. No control strategy compensates for this; therefore, we proposed a precision-focused corrective plan that targets between-month instability.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Glycated HemoglobinLaboratories, ClinicalHumansQuality ControlChromatography, High Pressure LiquidAccreditation

Stichwörter der Autor:innen

HbA1cQuality Goal IndexSix Sigma metricanalytical qualityexternal quality assessmentglycated hemoglobinimprecisioninternal quality controllaboratory accreditationtotal allowable error

Substanzen

Glycated Hemoglobinhemoglobin A1c protein, human

Autor:innen und Institutionen

  • Vedant Diagnostic Center, Rajkot, Gujarat, India.
  • Department of Laboratory Medicine, Vedant Diagnostic Center, Rajkot, Gujarat, India.

ORCID-Profile: Jagdish Chandarana, Kinjal Butani

Zitieren

APA

Chandarana, J., & Butani, K. (2026). Sigma metric and Quality Goal Index assessment of hemoglobin A1c analytical performance in an ambulatory accredited laboratory. Laboratory Medicine, 57(5). https://doi.org/10.1093/labmed/lmag052

Vancouver

Chandarana J, Butani K. Sigma metric and Quality Goal Index assessment of hemoglobin A1c analytical performance in an ambulatory accredited laboratory. Laboratory Medicine. 2026;57(5). doi:10.1093/labmed/lmag052

BibTeX
@article{Chandarana2026,
  author = {Chandarana, Jagdish and Butani, Kinjal},
  title = {Sigma metric and Quality Goal Index assessment of hemoglobin A1c analytical performance in an ambulatory accredited laboratory},
  journal = {Laboratory Medicine},
  year = {2026},
  volume = {57},
  number = {5},
  doi = {10.1093/labmed/lmag052}
}
Kostenlos starten →