A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report
Laboratory Medicine · 2026
DOI: 10.1093/labmed/lmaf061 ↗Kurz gefasst
Fallbericht in der Zeitschrift Laboratory Medicine, 2026. Verfasst von Nadia M Alqurini und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nephrology, University of Arkansas for Medical Sciences, Little Rock, AR, United States. Hauptthemen laut MEDLINE: Lactic Acid, 3-Hydroxybutyric Acid, Blood Gas Analysis. Die Arbeit nennt 7 Literaturangaben.
Abstract (Originalsprache)
Introduction
Laboratory tests play a crucial role in diagnosing and managing illness. Accurate interpretation, which includes evaluating false-negative and false-positive results, is essential for guiding appropriate clinical interventions and ensuring patient safety.
Methods
We present a case involving clinically significant discrepancies in lactate measurements in a patient with acute myeloid leukemia and diabetic ketoacidosis.
Results
These discrepancies were caused by interference from elevated levels of ketone bodies—specifically, β-hydroxybutyrate—which may have delayed recognition of the severity of the patient’s condition.
Discussion
Dilution and spike experiments suggested potential interferences in core laboratory instruments using the Trinder reaction, leading to inaccurate lactate readings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Nephrology, University of Arkansas for Medical Sciences, Little Rock, AR, United States.
- Department of Pathology, University of Arkansas for Medical Sciences, Little Rock, AR, United States.
ORCID-Profile: Nadia M Alqurini, Hoda A Hagrass
Zitieren
APA
Alqurini, N. M., Adhikari, B., & Hagrass, H. A. (2026). A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report. Laboratory Medicine. https://doi.org/10.1093/labmed/lmaf061
Vancouver
Alqurini NM, Adhikari B, Hagrass HA. A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report. Laboratory Medicine. 2026. doi:10.1093/labmed/lmaf061
BibTeX
@article{Alqurini2026,
author = {Alqurini, Nadia M and Adhikari, Bhawana and Hagrass, Hoda A},
title = {A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report},
journal = {Laboratory Medicine},
year = {2026},
doi = {10.1093/labmed/lmaf061}
}
EUVIMED