EUVIMEDEUVIMED entdecken
Fallbericht

A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report

Nadia M Alqurini iD, Bhawana Adhikari, Hoda A Hagrass iD

Laboratory Medicine · 2026

DOI: 10.1093/labmed/lmaf061 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Laboratory Medicine, 2026. Verfasst von Nadia M Alqurini und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nephrology, University of Arkansas for Medical Sciences, Little Rock, AR, United States. Hauptthemen laut MEDLINE: Lactic Acid, 3-Hydroxybutyric Acid, Blood Gas Analysis. Die Arbeit nennt 7 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Laboratory tests play a crucial role in diagnosing and managing illness. Accurate interpretation, which includes evaluating false-negative and false-positive results, is essential for guiding appropriate clinical interventions and ensuring patient safety.

Methods

We present a case involving clinically significant discrepancies in lactate measurements in a patient with acute myeloid leukemia and diabetic ketoacidosis.

Results

These discrepancies were caused by interference from elevated levels of ketone bodies—specifically, β-hydroxybutyrate—which may have delayed recognition of the severity of the patient’s condition.

Discussion

Dilution and spike experiments suggested potential interferences in core laboratory instruments using the Trinder reaction, leading to inaccurate lactate readings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lactic Acid3-Hydroxybutyric AcidBlood Gas AnalysisHumansDiabetic KetoacidosisFalse Positive ReactionsMaleLeukemia, Myeloid, AcuteMiddle AgedFemale

Stichwörter der Autor:innen

Trinderacidosisassay interferencebeta-hydroxbutyratecritical careenzymatic assaylactatelactic acidosis

Substanzen

Lactic Acid3-Hydroxybutyric Acid

Autor:innen und Institutionen

  • Department of Nephrology, University of Arkansas for Medical Sciences, Little Rock, AR, United States.
  • Department of Pathology, University of Arkansas for Medical Sciences, Little Rock, AR, United States.

ORCID-Profile: Nadia M Alqurini, Hoda A Hagrass

Zitieren

APA

Alqurini, N. M., Adhikari, B., & Hagrass, H. A. (2026). A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report. Laboratory Medicine. https://doi.org/10.1093/labmed/lmaf061

Vancouver

Alqurini NM, Adhikari B, Hagrass HA. A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report. Laboratory Medicine. 2026. doi:10.1093/labmed/lmaf061

BibTeX
@article{Alqurini2026,
  author = {Alqurini, Nadia M and Adhikari, Bhawana and Hagrass, Hoda A},
  title = {A falsely low lactate: discrepancies in lactate measurement between blood gas and core laboratory instruments due to β-hydroxybutyrate interference—a case report},
  journal = {Laboratory Medicine},
  year = {2026},
  doi = {10.1093/labmed/lmaf061}
}
Kostenlos starten →