EUVIMEDEUVIMED entdecken
Crossref

Detection of decreased granules in neutrophils by automated hematology analyzers XR-1000 and UniCel DxH 800

Yosuke Kato, Daisuke Sakamoto, Hiroaki Ohnishi, Tomohiko Taki

Laboratory Medicine · 2024

DOI: 10.1093/labmed/lmae047 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Laboratory Medicine, 2024. Verfasst von Yosuke Kato und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Technology, Kyorin University Faculty of Health Sciences, Mitaka, Japan. Hauptthemen laut MEDLINE: Neutrophils, Myelodysplastic Syndromes. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 29 Literaturangaben. Gefördert von: Sysmex Corporation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aimed to investigate the utility of neutrophil-related cell population data obtained by automated hematology analyzers in assessing myelodysplastic syndrome cases with decreased granules in neutrophils.

Methods

A total of 108 subjects were classified into normal granule (n = 35), hypogranulation (n = 37), or hypergranulation (n = 36) groups. Neutrophil cell area and granule area were measured by ImageJ. All samples were analyzed on the XR-1000 and UniCel DxH 800, and neutrophil-related parameters were compared among the 3 groups.

Results

Neutrophil cell area and the ratio of the granular area showed significant differences among the 3 groups; they were the highest in the hypergranulation group and lowest in the hypogranulation group. XR-1000 data showed significant differences in NE-SFL and NE-FSC among the 3 groups (P < .0001). NE-SFL and NE-FSC discriminated most accurately hypogranulation group against other groups. UniCel DxH 800 data showed significant differences in MN-V-NE, MN-MALS-N, MN-UMALS-NE, SD-UMALS-NE (P <.01), MN-LMALS-NE, and SD-LMALS-NE (P <.05) among the 3 groups. The combination of SD-V-NE and SD-LMALS-NE discriminated most accurately the hypogranulation group against the other groups.

Conclusion

NE-SFL and NE-FSC and the combination of SD-V-NE and SD-LMALS-NE are useful in detecting cases with decreased granules in neutrophils.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

NeutrophilsMyelodysplastic SyndromesHumansCytoplasmic GranulesFemaleMaleMiddle AgedAgedAutomation, LaboratoryAdultAged, 80 and over

Stichwörter der Autor:innen

UniCel DxH 800XR-1000cell population dataflow cytometryhypogranulationmyelodysplastic syndromes

Autor:innen und Institutionen

  • Department of Medical Technology, Kyorin University Faculty of Health Sciences, Mitaka, Japan.
  • Department of Laboratory Medicine, Kyorin University Faculty of Medicine, Mitaka, Japan.

Förderung

  • Sysmex Corporation

Zitieren

APA

Kato, Y., Sakamoto, D., Ohnishi, H., & Taki, T. (2024). Detection of decreased granules in neutrophils by automated hematology analyzers XR-1000 and UniCel DxH 800. Laboratory Medicine. https://doi.org/10.1093/labmed/lmae047

Vancouver

Kato Y, Sakamoto D, Ohnishi H, Taki T. Detection of decreased granules in neutrophils by automated hematology analyzers XR-1000 and UniCel DxH 800. Laboratory Medicine. 2024. doi:10.1093/labmed/lmae047

BibTeX
@article{Kato2024,
  author = {Kato, Yosuke and Sakamoto, Daisuke and Ohnishi, Hiroaki and Taki, Tomohiko},
  title = {Detection of decreased granules in neutrophils by automated hematology analyzers XR-1000 and UniCel DxH 800},
  journal = {Laboratory Medicine},
  year = {2024},
  doi = {10.1093/labmed/lmae047}
}