A simple method to overcome paraproteinemic interferences in chemistry and immunoassays

Rajarshi Sarkar iD

Laboratory Medicine · 2024

DOI: 10.1093/labmed/lmae042 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Laboratory Medicine, 2024. Verfasst von Rajarshi Sarkar. Beteiligte Einrichtung: Drs. Tribedi & Roy Diagnostic Laboratory, Kolkata, India. Hauptthemen laut MEDLINE: Immunoassay, Paraproteinemias, Paraproteins. Die Arbeit nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Interferences on chemistry and immunoassay results due to paraproteinemia may lead to erroneous diagnoses and treatment. Such interferences are difficult to recognize and even more difficult to deal with. This report describes 1 such case where multiple measurands were affected and how the interferant was overcome. Case Report Paraproteins present in an immunoglobulin (Ig)G-lambda multiple myeloma specimen interfered with results of total bilirubin, direct bilirubin, inorganic phosphate, iron, ferritin, and total thyroxine measured on 3 platforms: AU5800, Alinity ci, and cobas pure. Repeat testing upon dilution with normal saline or deproteinization by polyethylene glycol precipitation gave unsatisfactory results on some or all the affected measurands. Repeat testing after dilution of the interferant serum with a healthy serum corrected the anomalous results for all the affected measurands.

Conclusion

Dilution of paraproteinemic serum with a healthy serum of known concentrations appears to be the most suitable method to negate the effects of paraproteinemic interferences.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ImmunoassayParaproteinemiasParaproteinsHumansBilirubinBlood Chemical AnalysisMultiple MyelomaThyroxine

Stichwörter der Autor:innen

IgG myelomadeproteinizationerroneous resultsparaprotein interferencereaction curvessample dilution

Substanzen

BilirubinParaproteinsThyroxine

Autor:innen und Institutionen

  • Drs. Tribedi & Roy Diagnostic Laboratory, Kolkata, India.

ORCID-Profile: Rajarshi Sarkar

Zitieren

APA

Sarkar, R. (2024). A simple method to overcome paraproteinemic interferences in chemistry and immunoassays. Laboratory Medicine. https://doi.org/10.1093/labmed/lmae042

Vancouver

Sarkar R. A simple method to overcome paraproteinemic interferences in chemistry and immunoassays. Laboratory Medicine. 2024. doi:10.1093/labmed/lmae042

BibTeX
@article{Sarkar2024,
  author = {Sarkar, Rajarshi},
  title = {A simple method to overcome paraproteinemic interferences in chemistry and immunoassays},
  journal = {Laboratory Medicine},
  year = {2024},
  doi = {10.1093/labmed/lmae042}
}
Kostenlos starten →