Verifying the nonreporting hemolysis index for potassium, phosphate, magnesium, AST, LDH, iron, CA 19-9, and vitamin D, using Beckman Coulter AU5800 and DxI800 automated analyzers

Samuel Sheerin

Laboratory Medicine · 2024

DOI: 10.1093/labmed/lmae027 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Laboratory Medicine, 2024. Verfasst von Samuel Sheerin. Beteiligte Einrichtung: Department of Clinical Biochemistry, North Bristol NHS Trust-Southmead Hospital, Bristol, UK. Hauptthemen laut MEDLINE: Hemolysis, Blood Chemical Analysis. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 8 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Hemolysis is a common reason for nonreporting results in biochemistry and is measured using the hemolysis index (HI), with nonreporting limits set for analytes by manufacturers.

Objective

To verify the nonreporting HI limit for potassium, phosphate, magnesium, aspartate aminotransferase (AST), lactate dehydrogenase (LDH), iron, CA19-9, and vitamin D on the Beckman Coulter AU5800/DxI800 analyzers.

Method

Hemolysate was created from EDTA-lined tubes of whole blood using an osmotic shock procedure. The hemolysate underwent serial dilutions with saline and was spiked in paired serum. The delta changes in HI and analyte concentration were measured, assessed using regression analysis, and compared against calculated reference change values.

Results

A linear relationship between increasing HI and increasing analyte concentration (R2 > 0.9) was observed for potassium (y = 0.8864x), phosphate (y = 0.1079x), magnesium (y = 0.0678x), AST (y = 29.035x), and LDH (y = 350x). Increasing HI values did not have a linear effect on iron (y = −0.2544x), CA19-9 (y = 2.7019x), or vitamin D (y = 8.036x) concentrations.

Conclusion

The results from this experiment support increasing the HI nonreporting limit to 100 mg/dL for potassium; 200 mg/dL for magnesium; and 300 mg/dL for phosphate, CA19-9, and vitamin D. The iron assay is not affected by hemolysis as high as 500 mg/dL. The current HI nonreporting limit of 50 mg/dL is appropriate for LDH.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemolysisBlood Chemical AnalysisOsmotic PressureLimit of DetectionPotassiumPhosphatesMagnesiumAntigens, Tumor-Associated, CarbohydrateAspartate AminotransferasesL-Lactate DehydrogenaseVitamin DIronHumans

Stichwörter der Autor:innen

Beckman Coulterbiochemistryhemolysis indexindiceslimitsnonreporting

Substanzen

PotassiumPhosphatesMagnesiumcarbohydrate antigen 199, humanAntigens, Tumor-Associated, CarbohydrateAspartate AminotransferasesL-Lactate DehydrogenaseVitamin DIron

Autor:innen und Institutionen

  • Department of Clinical Biochemistry, North Bristol NHS Trust-Southmead Hospital, Bristol, UK.

Zitieren

APA

Sheerin, S. (2024). Verifying the nonreporting hemolysis index for potassium, phosphate, magnesium, AST, LDH, iron, CA 19-9, and vitamin D, using Beckman Coulter AU5800 and DxI800 automated analyzers. Laboratory Medicine. https://doi.org/10.1093/labmed/lmae027

Vancouver

Sheerin S. Verifying the nonreporting hemolysis index for potassium, phosphate, magnesium, AST, LDH, iron, CA 19-9, and vitamin D, using Beckman Coulter AU5800 and DxI800 automated analyzers. Laboratory Medicine. 2024. doi:10.1093/labmed/lmae027

BibTeX
@article{Sheerin2024,
  author = {Sheerin, Samuel},
  title = {Verifying the nonreporting hemolysis index for potassium, phosphate, magnesium, AST, LDH, iron, CA 19-9, and vitamin D, using Beckman Coulter AU5800 and DxI800 automated analyzers},
  journal = {Laboratory Medicine},
  year = {2024},
  doi = {10.1093/labmed/lmae027}
}
Kostenlos starten →