Modeling of a tissue expander with a radiofrequency identification port in postmastectomy radiation therapy planning

Fumiyasu Matsubayashi, Taro Takahashi, Hikaru Miyauchi, Yasushi Ito, Arisa Harada, Yasuo Yoshioka

Journal of Radiation Research · 2024

DOI: 10.1093/jrr/rrae004 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Radiation Research, 2024. Verfasst von Fumiyasu Matsubayashi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Oncology Department, Cancer Institute Hospital, Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan. Hauptthemen laut MEDLINE: Mastectomy, Radiotherapy Planning, Computer-Assisted, Tissue Expansion Devices. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The purpose of this study was to evaluate the dose attenuation of Motiva Flora® (Flora, Establishment Labs, Alajuela, Costa Rica) tissue expander with a radiofrequency identification port locator and to develop a model for accurate postmastectomy radiation therapy planning. Dose attenuation was measured using an EBT3 film (Ashland, Bridgewater, NJ), and the optimal material and density assignment for the radiofrequency identification coil for dose calculation were investigated using the AcurosXB algorithm on the Eclipse (Varian Medical Systems, Palo Alto, CA) treatment planning system. Additionally, we performed in vivo dosimetry analysis using irradiation tangential to the Flora tissue expander to validate the modeling accuracy. Dose attenuations downstream of the Flora radiofrequency identification coil was 1.29% for a 6 MV X-ray and 0.99% for a 10 MV X-ray when the coil was placed perpendicular to the beam. The most suitable assignments for the material and density of the radiofrequency identification coil were aluminum and 2.27 g/cm3, respectively, even though the coil was actually made of copper. Gamma analysis of in vivo dosimetry with criteria of 3% and 2 mm did not fail in the coil region. Therefore, we conclude that the model is reasonable for clinical use.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MastectomyRadiotherapy Planning, Computer-AssistedTissue Expansion DevicesHumansFemaleRadiotherapy DosageBreast NeoplasmsRadiometryDose-Response Relationship, Radiation

Stichwörter der Autor:innen

modelingpostmastectomy radiation therapyradiofrequency identificationtissue expander

Autor:innen und Institutionen

  • Radiation Oncology Department, Cancer Institute Hospital, Japanese Foundation for Cancer Research, 3-8-31 Ariake, Koto-ku, Tokyo 135-8550, Japan.

Zitieren

APA

Matsubayashi, F., Takahashi, T., Miyauchi, H., Ito, Y., Harada, A., & Yoshioka, Y. (2024). Modeling of a tissue expander with a radiofrequency identification port in postmastectomy radiation therapy planning. Journal of Radiation Research. https://doi.org/10.1093/jrr/rrae004

Vancouver

Matsubayashi F, Takahashi T, Miyauchi H, Ito Y, Harada A, Yoshioka Y. Modeling of a tissue expander with a radiofrequency identification port in postmastectomy radiation therapy planning. Journal of Radiation Research. 2024. doi:10.1093/jrr/rrae004

BibTeX
@article{Matsubayashi2024,
  author = {Matsubayashi, Fumiyasu and Takahashi, Taro and Miyauchi, Hikaru and Ito, Yasushi and Harada, Arisa and Yoshioka, Yasuo},
  title = {Modeling of a tissue expander with a radiofrequency identification port in postmastectomy radiation therapy planning},
  journal = {Journal of Radiation Research},
  year = {2024},
  doi = {10.1093/jrr/rrae004}
}
Kostenlos starten →