The effect of patient’s body weight, infusion connection point, and infusion pump position on intravenous multi-infusion drug delivery at low infusion rates suitable for premature neonates
Journal of Pharmacy and Pharmacology · 2024
DOI: 10.1093/jpp/rgad108 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Pharmacy and Pharmacology, 2024. Verfasst von Kamelia Krysiak und 3 weiteren Autor:innen. Beteiligte Einrichtung: The Rotunda Hospital, Parnell Square, Dublin, D01 P5W9, Ireland. Hauptthemen laut MEDLINE: Infusion Pumps, Drug Delivery Systems. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: School of Pharmacy and Biomolecular Science, Rotunda Foundation, B. Braun Ireland Ltd.
Abstract (Originalsprache)
Introduction
Parenteral drug administration in the neonatal intensive care involves complex pharmacotherapy adjusted for the patient’s weight, fluid allowance, and complex multi-infusion systems.
Objectives
We investigated the delivery rate of a model drug through a multi-infusion system consisting of six intravenous infusions.
Methods
Delivery rate of the model drug was determined after infusion initiation and termination. Measurements were collected spectrophotometrically in real time. Time to drug delivery and the amount of drug delivered were measured. Key findings The longest time to drug delivery was observed for a 500 g neonate model with a distal infusion connection point and neutral pump position (337 ± 30 min, P < 0.001). The shortest time was observed for a 1000 g neonate model in the combination of proximal infusion connection point and neutral pump position (18 ± 12 min, P < 0.05). The expected 100% of the drug was delivered only in two combinations: 500 g and 1000 g neonate models, proximal infusion connection point and neutral pump position (100.4 ± 4.7%, P = 0.819 and 100.2 ± 2.7%, P = 0.874, respectively). While the least drug was delivered to a 500 g neonate model in the combination of distal infusion connection point and neutral pump position (27.5 ± 5.8%, P < 0.001).
Conclusions
Delayed drug delivery to premature neonates due to multi-infusion systems may compromise accurate drug administration and lead to dosing errors.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- The Rotunda Hospital, Parnell Square, Dublin, D01 P5W9, Ireland.
- School of Pharmacy and Biomolecular Sciences, Royal College of Surgeons in Ireland University of Medicine and Health Sciences, Dublin, D02 H903, Ireland.
- School of Medicine, Royal College of Surgeons in Ireland University of Medicine and Health Sciences, Dublin, D02 H903Ireland.
ORCID-Profile: Fiona O'Brien
Förderung
- School of Pharmacy and Biomolecular Science
- Rotunda Foundation
- B. Braun Ireland Ltd
Zitieren
APA
Krysiak, K., Cleary, B., McCallion, N., & O'Brien, F. (2024). The effect of patient’s body weight, infusion connection point, and infusion pump position on intravenous multi-infusion drug delivery at low infusion rates suitable for premature neonates. Journal of Pharmacy and Pharmacology. https://doi.org/10.1093/jpp/rgad108
Vancouver
Krysiak K, Cleary B, McCallion N, O'Brien F. The effect of patient’s body weight, infusion connection point, and infusion pump position on intravenous multi-infusion drug delivery at low infusion rates suitable for premature neonates. Journal of Pharmacy and Pharmacology. 2024. doi:10.1093/jpp/rgad108
BibTeX
@article{Krysiak2024,
author = {Krysiak, Kamelia and Cleary, Brian and McCallion, Naomi and O'Brien, Fiona},
title = {The effect of patient’s body weight, infusion connection point, and infusion pump position on intravenous multi-infusion drug delivery at low infusion rates suitable for premature neonates},
journal = {Journal of Pharmacy and Pharmacology},
year = {2024},
doi = {10.1093/jpp/rgad108}
}
EUVIMED