Feasibility of Capillary Self-Collected Blood Specimens for Routine Outpatient Laboratory Testing
The Journal of Applied Laboratory Medicine · 04.11.2026 · 11 (6) · 1546-1557
DOI: 10.1093/jalm/jfag075 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 28.05.2026. Verfasst von Chelsea B Swartchick und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, United States. Hauptthemen laut MEDLINE: Blood Specimen Collection. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Tasso, RedDrop, YourBio.
Abstract (Originalsprache)
Background
Capillary self-collection (CSC) devices allow patients to collect blood samples at home, potentially reducing outpatient phlebotomy visits. This study aimed to (a) assess the patient experience with multiple commercially available CSC devices; (b) determine which laboratory tests are commonly ordered together, which could reduce the need for appointments; and (c) evaluate the analytical performance of these tests using CSC samples.
Methods
User experience for 3 CSC devices was evaluated. Clinical feasibility was determined by comparing test results in paired sera collected by venipuncture (VP) and CSC devices. VP samples were centrifuged within 2 h and tested immediately. CSC sera were centrifuged and tested both immediately and after delayed processing to simulate shipping temperature extremes (−20°C or 40°C). Basic metabolic panel, lipid panel, thyroid function cascade, and prostate-specific antigen were evaluated. Differences between VP and CSC collections were characterized according to their statistical and clinical differences.
Results
Patients reported CSC devices were easy to use and painless. Clinically significant differences between VP and CSC sera processed immediately were limited to potassium and bicarbonate. Following delayed processing of CSC serum samples (48 h at room temperature), clinically significant differences in potassium, bicarbonate, and glucose were observed. Frozen samples could not be analyzed. A 48 h delay at 40°C caused clinically significant differences in all analytes except creatinine, lipid panel, prostate-specific antigen, and the thyroid cascade.
Conclusion
Potassium, bicarbonate, and glucose were not stable at room temperature in CSC sera. CSC serum specimens may be a viable option but require analyte-specific evaluation and consideration of transportation conditions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, United States.
ORCID-Profile: Chelsea B Swartchick, Darci R Block
Förderung
- Tasso
- RedDrop
- YourBio
Zitieren
APA
Swartchick, C. B., Gurudu, R., Lexvold, N. S., Lasho, M. A., Baumann, N. A., Block, D. R., Katzman, B. M., & Meeusen, J. W. (2026). Feasibility of Capillary Self-Collected Blood Specimens for Routine Outpatient Laboratory Testing. The Journal of Applied Laboratory Medicine, 11(6), 1546-1557. https://doi.org/10.1093/jalm/jfag075
Vancouver
Swartchick CB, Gurudu R, Lexvold NS, Lasho MA, Baumann NA, Block DR, et al. Feasibility of Capillary Self-Collected Blood Specimens for Routine Outpatient Laboratory Testing. The Journal of Applied Laboratory Medicine. 2026;11(6):1546-1557. doi:10.1093/jalm/jfag075
BibTeX
@article{Swartchick2026,
author = {Swartchick, Chelsea B and Gurudu, Rachana and Lexvold, Nathanael S and Lasho, Michael A and Baumann, Nikola A and Block, Darci R and Katzman, Brooke M and Meeusen, Jeffrey W},
title = {Feasibility of Capillary Self-Collected Blood Specimens for Routine Outpatient Laboratory Testing},
journal = {The Journal of Applied Laboratory Medicine},
year = {2026},
volume = {11},
number = {6},
pages = {1546--1557},
doi = {10.1093/jalm/jfag075}
}