Assessment of Instrument-Specific Variation between Measured and Calculated Bicarbonate
The Journal of Applied Laboratory Medicine · 05.05.2026 · 11 (3) · 483-492
DOI: 10.1093/jalm/jfaf201 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 07.01.2026. Verfasst von Nga Yeung Tang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pathology, University of Illinois Chicago, Chicago, IL, United States. Hauptthemen laut MEDLINE: Bicarbonates, Blood Gas Analysis, Acid-Base Imbalance. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Bicarbonate concentration is widely used to assess acid-base disorders in patients, and the concentration can be directly measured or calculated. When discrepant results between measured bicarbonate (mHCO3−) and calculated bicarbonate (cHCO3−) are observed, it can lead to confusion among clinicians and potential for misdiagnosis. Here we assessed the agreement between cHCO3− from the Radiometer blood gas analyzer and mHCO3− from 3 different chemistry instruments.
Methods
Three institutions that measure bicarbonate using chemistry analyzers from 3 different manufacturers and derive bicarbonate on Radiometer blood gas analyzers participated in this study. De-identified patient data include plasma mHCO3− and blood gas cHCO3− results (arterial and venous) collected within +/− 20 min. Correlations and biases were determined.
Results
Deming regression analysis showed good correlations between cHCO3− and mHCO3−. Bland–Altman analysis revealed the greatest bias between Radiometer and Abbott Architect (−2.63 mmol/L), followed by Siemens Advia (0.49 mmol/L) and Beckman AU680 (−0.45 mmol/L).
Conclusions
This is the first study to compare cHCO3− from the Radiometer blood gas analyzer against mHCO3− results from multiple chemistry analyzers. Our results suggest that the Radiometer blood gas analyzer, Siemens Advia, and Beckman AU680 agree well with each other. However, bicarbonate results may be negatively biased when measured on the Abbott Architect compared with the other methods.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Pathology, University of Illinois Chicago, Chicago, IL, United States.
- Department of Pathology, University of Michigan, Ann Arbor, MI, United States.
- Departments of Laboratory Medicine and Pathology, University of Washington, Seattle, WA, United States.
- Department of Pathology and Laboratory Medicine, Corewell Health, Royal Oak, MI, United States.
- Department of Pathology and Laboratory Medicine, Oakland University William Beaumont School of Medicine, Auburn Hills, MI, United States.
ORCID-Profile: Nga Yeung Tang, Carmen Gherasim, Hsuan-Chieh Liao
Zitieren
APA
Tang, N. Y., Gherasim, C., Schroeder, L., Liao, H. C., & Sun, Q. (2026). Assessment of Instrument-Specific Variation between Measured and Calculated Bicarbonate. The Journal of Applied Laboratory Medicine, 11(3), 483-492. https://doi.org/10.1093/jalm/jfaf201
Vancouver
Tang NY, Gherasim C, Schroeder L, Liao HC, Sun Q. Assessment of Instrument-Specific Variation between Measured and Calculated Bicarbonate. The Journal of Applied Laboratory Medicine. 2026;11(3):483-492. doi:10.1093/jalm/jfaf201
BibTeX
@article{Tang2026,
author = {Tang, Nga Yeung and Gherasim, Carmen and Schroeder, Lee and Liao, Hsuan-Chieh and Sun, Qian},
title = {Assessment of Instrument-Specific Variation between Measured and Calculated Bicarbonate},
journal = {The Journal of Applied Laboratory Medicine},
year = {2026},
volume = {11},
number = {3},
pages = {483--492},
doi = {10.1093/jalm/jfaf201}
}
EUVIMED