Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay

John W Pickering iD, Stephen du Toit iD, Stacey Cheer, Vanessa Buchan, Rory Miller iD, Laura R Joyce iD, Martin Than iD, Christopher M Florkowski iD

The Journal of Applied Laboratory Medicine · 05.03.2026 · 11 (2) · 295-306

DOI: 10.1093/jalm/jfaf181 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 12.12.2025. Verfasst von John W Pickering und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Emergency Medicine, Christchurch Hospital, Christchurch, New Zealand. Hauptthemen laut MEDLINE: Troponin I, Point-of-Care Systems. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Heart Foundation (1983), Lottery Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The QuidelOrtho TriageTrue® point-of-care (POC) high-sensitivity troponin I (hs-TnI) assay has been previously validated. We aimed to independently verify the assay’s analytical performance and concordance with other assays and to verify secondary POC devices.

Methods

Intra-assay precision: 5 whole blood samples spanning the measuring interval were analyzed up to 20 times in succession. Inter-analyzer and inter-lot precision studies were also carried out and a precision curve generated. Hemolysis was assessed by spiking 8 samples with concentrations from 2 to 320 ng/L with hemolysate. Concordance between >200 TriageTrue samples and one other hs-TnI POC and 4 laboratory analyzers was assessed by Pearson correlation and kappa statistics at the limit of quantitation and upper reference limit. Paired measurements from 9 samples between a primary verified device and 8 secondary devices were compared against prespecified difference limits.

Results

Intra-assay precision CVs were 25% and 12.2% at mean concentrations of 2.4 ng/L and 8.5 ng, respectively, and ranged from 2.8% to 15.9% at higher concentrations >10 ng/L. There was minimal interference by hemolysis up to 10 g/L, which exceeded the manufacturer’s claim of 1.96 g/L. Correlation coefficients were 0.89 to 0.99 with other assays with the exception of the Roche hs-TnT assay, where it was 0.79. These were similar to the laboratory assays. Of 116 comparisons with 8 secondary analyzers, all fell within acceptable limits.

Conclusions

The TriageTrue assay performed as reported in the package insert. The assay characteristics offer acceptable performance for use within the intended medical settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Troponin IPoint-of-Care SystemsHumansReproducibility of ResultsHemolysisLimit of Detection

Substanzen

Troponin I

Autor:innen und Institutionen

  • Department of Emergency Medicine, Christchurch Hospital, Christchurch, New Zealand.
  • Department of Medicine, Christchurch Heart Institute, University of Otago Christchurch, Christchurch, New Zealand.
  • Waikato Hospital Laboratory, Health NZ, Hamilton, New Zealand.
  • Canterbury Health Laboratories, Christchurch Hospital, Christchurch, New Zealand.
  • Department of General Practice and Rural Health, University of Otago, Dunedin, New Zealand.
  • Thames Hospital, Thames, New Zealand.

ORCID-Profile: John W Pickering, Stephen du Toit, Rory Miller, Laura R Joyce, Martin Than, Christopher M Florkowski

Förderung

  • Heart Foundation (1983)
  • Lottery Health

Zitieren

APA

Pickering, J. W., du Toit, S., Cheer, S., Buchan, V., Miller, R., Joyce, L. R., Than, M., & Florkowski, C. M. (2026). Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay. The Journal of Applied Laboratory Medicine, 11(2), 295-306. https://doi.org/10.1093/jalm/jfaf181

Vancouver

Pickering JW, du Toit S, Cheer S, Buchan V, Miller R, Joyce LR, et al. Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay. The Journal of Applied Laboratory Medicine. 2026;11(2):295-306. doi:10.1093/jalm/jfaf181

BibTeX
@article{Pickering2026,
  author = {Pickering, John W and du Toit, Stephen and Cheer, Stacey and Buchan, Vanessa and Miller, Rory and Joyce, Laura R and Than, Martin and Florkowski, Christopher M},
  title = {Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2026},
  volume = {11},
  number = {2},
  pages = {295--306},
  doi = {10.1093/jalm/jfaf181}
}
Kostenlos starten →