Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay
The Journal of Applied Laboratory Medicine · 05.03.2026 · 11 (2) · 295-306
DOI: 10.1093/jalm/jfaf181 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 12.12.2025. Verfasst von John W Pickering und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Emergency Medicine, Christchurch Hospital, Christchurch, New Zealand. Hauptthemen laut MEDLINE: Troponin I, Point-of-Care Systems. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Heart Foundation (1983), Lottery Health.
Abstract (Originalsprache)
Background
The QuidelOrtho TriageTrue® point-of-care (POC) high-sensitivity troponin I (hs-TnI) assay has been previously validated. We aimed to independently verify the assay’s analytical performance and concordance with other assays and to verify secondary POC devices.
Methods
Intra-assay precision: 5 whole blood samples spanning the measuring interval were analyzed up to 20 times in succession. Inter-analyzer and inter-lot precision studies were also carried out and a precision curve generated. Hemolysis was assessed by spiking 8 samples with concentrations from 2 to 320 ng/L with hemolysate. Concordance between >200 TriageTrue samples and one other hs-TnI POC and 4 laboratory analyzers was assessed by Pearson correlation and kappa statistics at the limit of quantitation and upper reference limit. Paired measurements from 9 samples between a primary verified device and 8 secondary devices were compared against prespecified difference limits.
Results
Intra-assay precision CVs were 25% and 12.2% at mean concentrations of 2.4 ng/L and 8.5 ng, respectively, and ranged from 2.8% to 15.9% at higher concentrations >10 ng/L. There was minimal interference by hemolysis up to 10 g/L, which exceeded the manufacturer’s claim of 1.96 g/L. Correlation coefficients were 0.89 to 0.99 with other assays with the exception of the Roche hs-TnT assay, where it was 0.79. These were similar to the laboratory assays. Of 116 comparisons with 8 secondary analyzers, all fell within acceptable limits.
Conclusions
The TriageTrue assay performed as reported in the package insert. The assay characteristics offer acceptable performance for use within the intended medical settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Emergency Medicine, Christchurch Hospital, Christchurch, New Zealand.
- Department of Medicine, Christchurch Heart Institute, University of Otago Christchurch, Christchurch, New Zealand.
- Waikato Hospital Laboratory, Health NZ, Hamilton, New Zealand.
- Canterbury Health Laboratories, Christchurch Hospital, Christchurch, New Zealand.
- Department of General Practice and Rural Health, University of Otago, Dunedin, New Zealand.
- Thames Hospital, Thames, New Zealand.
ORCID-Profile: John W Pickering, Stephen du Toit, Rory Miller, Laura R Joyce, Martin Than, Christopher M Florkowski
Förderung
- Heart Foundation (1983)
- Lottery Health
Zitieren
APA
Pickering, J. W., du Toit, S., Cheer, S., Buchan, V., Miller, R., Joyce, L. R., Than, M., & Florkowski, C. M. (2026). Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay. The Journal of Applied Laboratory Medicine, 11(2), 295-306. https://doi.org/10.1093/jalm/jfaf181
Vancouver
Pickering JW, du Toit S, Cheer S, Buchan V, Miller R, Joyce LR, et al. Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay. The Journal of Applied Laboratory Medicine. 2026;11(2):295-306. doi:10.1093/jalm/jfaf181
BibTeX
@article{Pickering2026,
author = {Pickering, John W and du Toit, Stephen and Cheer, Stacey and Buchan, Vanessa and Miller, Rory and Joyce, Laura R and Than, Martin and Florkowski, Christopher M},
title = {Analytical Verification of a Point-of-Care High-Sensitivity Troponin I Assay},
journal = {The Journal of Applied Laboratory Medicine},
year = {2026},
volume = {11},
number = {2},
pages = {295--306},
doi = {10.1093/jalm/jfaf181}
}
EUVIMED