EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

Method Validation of Dipeptidyl Peptidase 3 Assay

Leslie J Donato iD, Jeffrey W Meeusen, Renee' J Scott, Vlad C Vasile, Allan S Jaffe iD

The Journal of Applied Laboratory Medicine · 05.03.2026 · 11 (2) · 350-356

DOI: 10.1093/jalm/jfaf173 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 13.11.2025. Verfasst von Leslie J Donato und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine and Pathology, Mayo Clinic College of Medicine and Science, Rochester, MN, United States. Hauptthemen laut MEDLINE: Dipeptidyl-Peptidases and Tripeptidyl-Peptidases, Luminescent Measurements. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Dipeptidyl peptidase 3 (DPP3) is a peptidase released from dying cells. It cleaves proteins in the renin–angiotensin pathway, which can result in hemodynamic instability. At elevated concentrations DPP3 is associated with worse outcomes, particularly in patients with shock. Herein we describe the assay performance of a DPP3 assay (4TEEN4 Pharmaceuticals GmbH) in human plasma.

Methods

DPP3 concentration was measured using the DPP3 immunoluminometric assay (4TEEN4 Pharmaceuticals GmbH) and the signal was read using a luminometer (Berthold Centro LB963). Analytical performance was established for precision, linearity, accuracy, detection limit, analytical specificity, reference interval, kit lot-to-lot comparison, specimen type, and sample stability.

Results

Limit of detection was verified at 1.6 ng/mL in EDTA plasma with a coefficient of variation (CV) of <10%. Precision studies revealed a CV ≤ 6% at 28.5 ng/mL and 59.4 ng/mL and comparability with a manufacturer performed assay was demonstrated between 7.7 and 195.2 ng/mL. An upper 97.5% limit of 22 ng/mL without age or sex associations was verified in healthy donors. The assay was not susceptible to interference from lipemia or bilirubin. However, measured DPP3 concentrations increased linearly with increasing hemolysis. DPP3 concentrations are stable in EDTA plasma for up to 24 h and at least 11 months when stored ambient or at −80°C, respectively.

Conclusions

DPP3 can be measured precisely in EDTA plasma using the immunoluminometric DPP3 assay. Given the potential clinical use of DPP3 in critical care patients, caution should be taken to avoid inducing pre-analytical hemolysis during sample collection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dipeptidyl-Peptidases and Tripeptidyl-PeptidasesLuminescent MeasurementsHumansReproducibility of ResultsMaleFemaleLimit of DetectionAdultMiddle Aged

Substanzen

Dipeptidyl-Peptidases and Tripeptidyl-PeptidasesDPP3 protein, human

Autor:innen und Institutionen

  • Department of Laboratory Medicine and Pathology, Mayo Clinic College of Medicine and Science, Rochester, MN, United States.
  • Division of Cardiology, Mayo Clinic, Rochester, MN, United States.

ORCID-Profile: Leslie J Donato, Allan S Jaffe

Zitieren

APA

Donato, L. J., Meeusen, J. W., Scott, R. J., Vasile, V. C., & Jaffe, A. S. (2026). Method Validation of Dipeptidyl Peptidase 3 Assay. The Journal of Applied Laboratory Medicine, 11(2), 350-356. https://doi.org/10.1093/jalm/jfaf173

Vancouver

Donato LJ, Meeusen JW, Scott RJ, Vasile VC, Jaffe AS. Method Validation of Dipeptidyl Peptidase 3 Assay. The Journal of Applied Laboratory Medicine. 2026;11(2):350-356. doi:10.1093/jalm/jfaf173

BibTeX
@article{Donato2026,
  author = {Donato, Leslie J and Meeusen, Jeffrey W and Scott, Renee' J and Vasile, Vlad C and Jaffe, Allan S},
  title = {Method Validation of Dipeptidyl Peptidase 3 Assay},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2026},
  volume = {11},
  number = {2},
  pages = {350--356},
  doi = {10.1093/jalm/jfaf173}
}