EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

The Preanalytical Process Matters: Impact of Blood Collection Tubes on the Measurement of Vitamin D Using Liquid Chromatography Tandem Mass Spectrometry

Sae Ochi iD, Akihiro Kunisawa, Tomokazu Matsuura, Hiroyasu Miyamoto, Yuichi Ikeda, Masaki Miyasaka, Ayasa Tajima, Reiko Kisugi, Daisuke Kawakami, Tsuyoshi Nakanishi, Nobuhiro Hanafusa, Yutaka Furutani

The Journal of Applied Laboratory Medicine · 04.11.2025 · 10 (6) · 1504-1514

DOI: 10.1093/jalm/jfaf141 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, online erschienen am 07.10.2025. Verfasst von Sae Ochi und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, The Jikei University School of Medicine, Tokyo, Japan. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Vitamin D, Blood Specimen Collection. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Shimadzu Corp, Sysmex, TOSO.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Liquid chromatography tandem mass spectrometry (LC-MS/MS) is a highly sensitive analytical tool that has become the gold standard for measuring the concentrations of some steroidal hormones and therapeutic drugs in blood. However, the application of this technology in different clinical laboratories faces challenges related to the preanalytical process. In this study, we investigated how the materials used in blood collection tubes affect the measurement of serum 25-hydroxyvitamin D [25(OH)D] by LC-MS/MS.

Methods

Serum samples were collected in 20 different blood collection tubes made by various manufacturers. The samples were processed under controlled conditions to simulate clinical practice, followed by measurement using an LC-MS/MS system. The data were compared with measurements obtained from a chemiluminescent enzyme immunoassay.

Results

Significant variability in measurement outcomes was observed depending on the type of blood collection tube used. Specific blood collection tubes from a manufacturer that employed specific separator gels exhibited interference, as indicated by broad peaks in chromatograms. This interference complicated the quantification of 25(OH)D. The interference was consistent across multiple samples, indicating a systemic problem related to the materials used in the blood collection tubes.

Conclusions

These findings highlight the need to evaluate blood collection tubes from various manufacturers prior to the clinical implementation of LC-MS/MS measurements. Because LC-MS/MS has unique potential error sources, our results indicate the importance in utilizing LC-MS/MS international standards to ensure accurate and reliable results in clinical laboratories.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tandem Mass SpectrometryVitamin DBlood Specimen CollectionHumansChromatography, LiquidReproducibility of Results

Substanzen

Vitamin D25-hydroxyvitamin D

Autor:innen und Institutionen

  • Department of Laboratory Medicine, The Jikei University School of Medicine, Tokyo, Japan.
  • Analytical Business Unit, Shimadzu Europa GmbH, Duisburg, Germany.
  • Department of Central Clinical Laboratory, The Jikei University Daisan Hospital, Tokyo, Japan.
  • Department of Central Clinical Laboratory, The Jikei University Hospital, Tokyo, Japan.
  • Analytical & Measuring Instruments Division, Shimadzu Corporation, Kyoto, Japan.

ORCID-Profile: Sae Ochi

Förderung

  • Shimadzu Corp
  • Sysmex
  • TOSO

Zitieren

APA

Ochi, S., Kunisawa, A., Matsuura, T., Miyamoto, H., Ikeda, Y., Miyasaka, M., Tajima, A., Kisugi, R., Kawakami, D., Nakanishi, T., Hanafusa, N., & Furutani, Y. (2025). The Preanalytical Process Matters: Impact of Blood Collection Tubes on the Measurement of Vitamin D Using Liquid Chromatography Tandem Mass Spectrometry. The Journal of Applied Laboratory Medicine, 10(6), 1504-1514. https://doi.org/10.1093/jalm/jfaf141

Vancouver

Ochi S, Kunisawa A, Matsuura T, Miyamoto H, Ikeda Y, Miyasaka M, et al. The Preanalytical Process Matters: Impact of Blood Collection Tubes on the Measurement of Vitamin D Using Liquid Chromatography Tandem Mass Spectrometry. The Journal of Applied Laboratory Medicine. 2025;10(6):1504-1514. doi:10.1093/jalm/jfaf141

BibTeX
@article{Ochi2025,
  author = {Ochi, Sae and Kunisawa, Akihiro and Matsuura, Tomokazu and Miyamoto, Hiroyasu and Ikeda, Yuichi and Miyasaka, Masaki and Tajima, Ayasa and Kisugi, Reiko and Kawakami, Daisuke and Nakanishi, Tsuyoshi and Hanafusa, Nobuhiro and Furutani, Yutaka},
  title = {The Preanalytical Process Matters: Impact of Blood Collection Tubes on the Measurement of Vitamin D Using Liquid Chromatography Tandem Mass Spectrometry},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2025},
  volume = {10},
  number = {6},
  pages = {1504--1514},
  doi = {10.1093/jalm/jfaf141}
}
Kostenlos starten →