EUVIMEDEUVIMED entdecken
Crossref

Using Aggregated Proficiency Testing Results to Identify Systematic Error

Uzay Kırbıyık, J Rex Astles iD

The Journal of Applied Laboratory Medicine · 2025

DOI: 10.1093/jalm/jfaf126 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Applied Laboratory Medicine, 2025. Verfasst von Uzay Kırbıyık und 1 weiteren Autor:innen. Beteiligte Einrichtung: Division of Laboratory Systems, OLSR (Office of Laboratory Systems and Response), Centers for Disease Control and Prevention (CDC), Atlanta, GA, Unit…. Hauptthemen laut MEDLINE: Laboratory Proficiency Testing, Laboratories, Clinical, Clinical Laboratory Techniques. Die Arbeit nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Proficiency testing (PT) should identify systematic errors, which are likely to recur. The Clinical Laboratory Improvement Amendments of 1988 (CLIA) acceptance limits (ALs) include 3 standard deviations (3SD) and concentration limits. We investigated the ability of PT to detect systematic error, especially as affected by the different AL types.

Methods

We removed any ungradable, duplicate, and irregular scores from CLIA laboratory PT data from 2008 to 2018. We calculated the overall miss rate, unsatisfactory event rate, i.e., score <80 (4 of 5 correct), and event rates for score 100 to 0. We used paired t-tests and the Wilcoxon signed-rank test to compare miss rates and unsatisfactory event rates between short- and long-term PT participants. We used the binomial distribution to estimate the expected event scores under the assumption that all misses were independent (random). We compared observed event scores with their expected values as a ratio.

Results

Forty thousand five hundred ninety-six laboratories produced 15 140 128 event scores for 75 analytes. The distribution of event scores was skewed toward multiple event misses (score 0–60) compared to the predicted distribution. Miss rates and unsatisfactory rates were significantly higher for short-term laboratories. Plotting the log ratio of observed vs expected rates for event scores showed that the degree of systematic effect was substantial. The magnitude of the effect was less for 3SD ALs.

Conclusions

In an event, PT misses are often dependent. All ALs detected systematic error. Expressing systematic error using PT data could help to identify and remediate analytical issues.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Laboratory Proficiency TestingLaboratories, ClinicalClinical Laboratory TechniquesHumans

Autor:innen und Institutionen

  • Division of Laboratory Systems, OLSR (Office of Laboratory Systems and Response), Centers for Disease Control and Prevention (CDC), Atlanta, GA, United States.

ORCID-Profile: J Rex Astles

Zitieren

APA

Kırbıyık, U., & Astles, J. R. (2025). Using Aggregated Proficiency Testing Results to Identify Systematic Error. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfaf126

Vancouver

Kırbıyık U, Astles JR. Using Aggregated Proficiency Testing Results to Identify Systematic Error. The Journal of Applied Laboratory Medicine. 2025. doi:10.1093/jalm/jfaf126

BibTeX
@article{Krbyk2025,
  author = {Kırbıyık, Uzay and Astles, J Rex},
  title = {Using Aggregated Proficiency Testing Results to Identify Systematic Error},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2025},
  doi = {10.1093/jalm/jfaf126}
}
Kostenlos starten →